PEDIATRIC-PATIENT EXPOSURES FROM CT EXAMINATIONS - GE CT/T 9800 SCANNER

被引:32
作者
FEARON, T
VUCICH, J
机构
[1] CHILDRENS HOSP, NATL MED CTR, DEPT RADIOL, 111 MICHIGAN AVE NW, WASHINGTON, DC 20010 USA
[2] GEORGE WASHINGTON UNIV, SCH MED, DEPT RADIOL, WASHINGTON, DC 20037 USA
[3] GEORGE WASHINGTON UNIV, SCH MED, DEPT CHILD HLTH & DEV, WASHINGTON, DC 20037 USA
[4] NIH, DEPT RADIOL, BETHESDA, MD 20205 USA
关键词
D O I
10.2214/ajr.144.4.805
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Computed Tomography Dose Index (CTDI) values for typical CT examinations of children for a GE CT/T 9800 scanner are presented and they are compared with measured entrance skin absorbed doses of pediatric patients under clinical situations. Pediatric entrance skin absorbed doses were 1.1-2.4 rad (cGy [centrigrays]) for chest and abdomen examinations, 2.0-3.4 rad (cGy) for pediatric head examinations and 3.2-4.2 rad (cGy) for infant (.ltoreq. 6 mo.) head examinations. CTDI measurements in a cylindrical Lucite head phantom predicted typical absorbed doses to within .apprx. 5% for chest and abdomen examinations and to within .apprx. 15% for head examinations, when corrections for amperage differences are taken into account.