EFFECTS OF IMMERSION IN CHOLESTEROL-LIPID SOLUTION ON THE TENSILE AND FATIGUE PROPERTIES OF ELASTOMERIC POLYMERS FOR BLOOD PUMP APPLICATIONS

被引:14
作者
HAYASHI, K [1 ]
MATSUDA, T [1 ]
TAKANO, H [1 ]
UMEZU, M [1 ]
机构
[1] NATL CARDIOVASC CTR,RES INST,DEPT ARTIFICIAL ORGANS,SUITA,OSAKA 565,JAPAN
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1984年 / 18卷 / 08期
关键词
BIOLOGICAL MATERIALS - Blood - BIOMEDICAL EQUIPMENT - Pumps;
D O I
10.1002/jbm.820180808
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Tensile and fatigue properties were studied on three kinds of segmented polyether polyurethanes developed for blood pump applications. To study the effects of plasma constitutents, specimens were immersed in a cholesterol-lipid solution whose composition is similar to that of plasma. The results obtained were compared with those observed in a saline solution which had been reported previously. Immersion in the cholesterol-lipid solution has influence on the mechanical properties of Toyobo TM% polyurethane and Avcothane 51: Its effect is enhanced by cyclic deformation. Cholesterol-lipids change the mechanical characteristics of Avcothane 51 more remarkably than those of Toyobo TM5 polyurethane possibly because the former contains about 10% silicone. On the other hand, there is little effect of the constituents of the cholesterol-lipid solution on the static and dynamic mechanical properties of Biomer, although moisture and cyclic deformation have some influence on the mechanical characteristics.
引用
收藏
页码:939 / 951
页数:13
相关论文
共 18 条
  • [1] Black J., 1981, BIOL PERFORMANCE MAT
  • [2] SURFACE AND BULK CHARACTERISTICS OF A POLYETHER URETHANE FOR ARTIFICIAL HEARTS
    BORETOS, JW
    PIERCE, WS
    BAIER, RE
    LEROY, AF
    DONACHY, HJ
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1975, 9 (03): : 327 - 340
  • [3] LIPID ABSORPTION BY SILICONE-RUBBER HEART VALVE POPPETS IN-VIVO AND IN-VITRO RESULTS
    CARMEN, R
    MUTHA, SC
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1972, 6 (05): : 327 - &
  • [4] DESIGN OF PUSHER-PLATE-TYPE LEFT-VENTRICULAR ASSIST DEVICE BASED ON MECHANICAL ANALYSES
    HAYASHI, K
    NAKAMURA, T
    TAKANO, H
    UMEZU, M
    TAENAKA, Y
    MATSUDA, T
    [J]. ARTIFICIAL ORGANS, 1984, 8 (02) : 204 - 214
  • [5] HAYASHI K, INTRO BIOMATERIALS C
  • [6] HAYASHI K, UNPUB J BIOMED MAT R
  • [7] KIRALY RJ, 1977, T AM SOC ART INT ORG, V23, P127
  • [8] IMPROVING HEMATOLOGIC PICTURE IN LONG-TERM SURVIVING CALVES WITH TOTAL ARTIFICIAL HEARTS
    KOLFF, J
    HERSHGOLD, EJ
    HADFIELD, C
    OLSEN, DB
    LAWSON, J
    KOLFF, WJ
    [J]. ARTIFICIAL ORGANS, 1979, 3 (01) : 97 - 103
  • [9] TIME-DEPENDENT FAILURE OF A POLYOLEFIN RUBBER CANDIDATE MATERIAL FOR BLOOD PUMP APPLICATIONS
    MCKENNA, GB
    PENN, RW
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1980, 14 (05): : 689 - 703
  • [10] PHYSICAL TESTING OF ELASTOMERS FOR CARDIOVASCULAR APPLICATIONS
    MCMILLIN, CR
    [J]. ARTIFICIAL ORGANS, 1983, 7 (01) : 78 - 91