INVITRO VALIDATION OF A THERMODILUTION RIGHT VENTRICULAR EJECTION FRACTION METHOD

被引:22
作者
FERRIS, SE [1 ]
KONNO, M [1 ]
机构
[1] BAXTER HLTH CARE CORP,EDWARDS CRIT CARE DIV,SANTA ANA,CA
来源
JOURNAL OF CLINICAL MONITORING | 1992年 / 8卷 / 01期
关键词
MEASUREMENT TECHNIQUES; THERMODILUTION; HEART; VENTRICLE; RIGHT; EJECTION FRACTION;
D O I
10.1007/BF01618092
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A thermodilution catheter and computer system has been developed to measure right ventricular ejection fraction and volumes. To evaluate the performance of this method, the thermodilution system was evaluated in an in vitro pulsatile flow model. Thermodilution measurements of ejection fraction (EF), cardiac output (CO), stroke volume (SV), end-diastolic volume (EDV), and end-systolic volume (ESV) were compared with known values in a pulsatile flow bench. Thermodilution EF measurements correlated very well with the pulsatile flow model (r2 = 0.95, m [slope] = 0.85, SEE = 4.0 EFU). Thermodilution CO and SV were highly predictive of actual pulsatile flow (r2 = 0.99, m = 0.99, SEE = 187 ml/min and r2 = 0.98, m = 0.96, SEE = 2.5 ml, respectively). Thermodilution end-diastolic and end-systolic volume measurements resulted in low mean eror, -1.8% and 0-6%, respectively. The standard deviations of the error for EDV and ESV were 11.0% and 16.4%. The thermodilution measurements were repeatable, with CO, SV, and EF coefficients of variation of 3.2%, 3.3%, and 4.7%, respectively. EDV and ESV were slightly more variable, with coefficients of variation of 5.5% and 7.2%, respectively.
引用
收藏
页码:74 / 80
页数:7
相关论文
共 13 条
  • [1] Voelker W., Gruber H.P., Ickrath O., Et al., Determination of right ventricular ejection fraction by thermodilution technique—a comparison to biplane cineventriculography, Intensive Care Med, 14, pp. 461-466, (1988)
  • [2] Hurford W., The right ventricle and critical illness: a review of anatomy, physiology, and clinical evaluation of its function, Intensive Care Med, 14, pp. 448-457, (1988)
  • [3] Dhainaut J.F., Right ventricular dysfunction in patients with septic shock, Intensive Care Med, 14, pp. 488-491, (1988)
  • [4] Boldt J., Revascularization of the right coronary artery: influence on thermodilution right ventricular ejection fraction, J. Cardiothorac Anesth, 2, pp. 140-146, (1988)
  • [5] Kay H.R., Manoucher A., Barash P., Et al., Measurement of ejection fraction by thermal dilution techniques, J Surg Res, 34, pp. 337-346, (1983)
  • [6] Vincent J., Thermodilution measurement of right ventricular ejection fraction with a modified pulmonary artery catheter, Intensive Care Med, 12, pp. 33-38, (1986)
  • [7] Dhainaut J., Bedside evaluation of right ventricular performance using a rapid computerized thermodilution method, Crit Care Med, 15, pp. 148-152, (1987)
  • [8] Eaton S., Preliminary pulsatile flow bench validation of a thermodilution right ventricular ejection fraction system, Intensive Care Med, 14, pp. 458-460, (1988)
  • [9] Daskalov I.K., Gueorguiev L.V., Matveev M.G., Hybrid computer for dye-dilution curves, Med Biol Eng Comput, 68, pp. 68-72, (1978)
  • [10] Ganz W., Swan J.C., Measurement of blood flow by thermodilution, Am J Cardiol, 29, pp. 241-246, (1972)