COMPARISON OF A MODIFIED THIAZOLE ORANGE TECHNIQUE WITH A FULLY AUTOMATED ANALYZER FOR RETICULOCYTE COUNTING

被引:38
作者
BOWEN, D [1 ]
BENTLEY, N [1 ]
HOY, T [1 ]
CAVILL, I [1 ]
机构
[1] UNIV COLL CARDIFF,COLL MED,CARDIFF CF4 4XN,WALES
关键词
D O I
10.1136/jcp.44.2.130
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Two independent methods for quantitating reticulocyte counts were compared. One used a modified thiazole orange technique and a flow cytometer (Becton Dickinson FACS); the other was a fully automated whole blood analyser (Sysmex R1000). Both methods gave comparable results with a coefficient of variation of less than 5%. Samples measured using the R1000 showed a negligible decrease in the reticulocyte count over five days at room temperature, although there was evidence of continuing intracellular maturation: with thiazole orange there was an apparent increase. A practical reference range of 20-70 x 10(9)/1 was established from 89 normal subjects. The close correlation between the two independent estimates indicates the validity of the quantitation of the reticulocyte count and shows that automation allows significant changes within and below the normal range to be detected with a degree of reliability which was not previously possible.
引用
收藏
页码:130 / 133
页数:4
相关论文
共 8 条
  • [1] COUNTING RETICULOCYTES BY FLOW-CYTOMETRY - USE OF THIAZOLE ORANGE
    CARTER, JM
    MCSWEENEY, PA
    WAKEM, PJ
    NEMET, AM
    [J]. CLINICAL AND LABORATORY HAEMATOLOGY, 1989, 11 (03): : 267 - 271
  • [2] EVALUATION OF RETICULOCYTE COUNTS BY FLOW-CYTOMETRY IN A ROUTINE LABORATORY
    FERGUSON, DJ
    LEE, SF
    GORDON, PA
    [J]. AMERICAN JOURNAL OF HEMATOLOGY, 1990, 33 (01) : 13 - 17
  • [4] AN AUTOMATED OPTOELECTRONIC RETICULOCYTE COUNTER
    KOJIMA, K
    NIRI, M
    SETOGUCHI, K
    TSUDA, I
    TATSUMI, N
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 92 (01) : 57 - 61
  • [5] THIAZOLE ORANGE - A NEW DYE FOR RETICULOCYTE ANALYSIS
    LEE, LG
    CHEN, CH
    CHIU, LA
    [J]. CYTOMETRY, 1986, 7 (06): : 508 - 517
  • [6] TANKE HJ, 1983, BLOOD, V61, P109
  • [7] TICHELLI A, 1990, AM J CLIN PATHOL, V11, P267
  • [8] VERWILGEN RL, 1988, QUALITY ASSURANCE HA, P33