NONINVASIVE BLOOD-PRESSURE MEASUREMENTS WITH SINGLE AND TWIN-HOSE SYSTEMS - DO MIXTURES MATTER

被引:2
作者
AMOORE, JN
SCOTT, DHT
机构
[1] Department of Medical Physics and Medical Engineering, Royal Infirmary of Edinburgh
[2] Royal Infirmary of Edinburgh, Edinburgh
关键词
MEASUREMENT TECHNIQUES; BLOOD PRESSURE;
D O I
10.1111/j.1365-2044.1993.tb07596.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Noninvasive blood pressure monitors which use twin‐ and single‐hose cuffs complicate the movement of patients between bed spaces with different cuff‐hose systems. We assessed, using a commercially available noninvasive blood pressure monitor function analyser, the use of single‐hose cuffs coupled via a Y‐piece adaptor to the twin‐hose of noninvasive monitors designed for use with twin‐hose cuffs. Assessments were made at three simulated pressures (200/150 mmHg, 120/80 mmHg, and 60/30 mmHg) and at normal and weak arterial pulsations. The use of a single‐hose cuff attached with the Y‐piece adaptor did not adversely affect the performance of the monitors. Copyright © 1993, Wiley Blackwell. All rights reserved
引用
收藏
页码:799 / 802
页数:4
相关论文
共 6 条
[1]   ASSESSMENT OF OSCILLOMETRIC NONINVASIVE BLOOD-PRESSURE MONITORS USING THE DYNATECH NEVADA CUFFLINK ANALYZER [J].
AMOORE, JN .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1993, 17 (01) :25-31
[2]   CARDIAC CYCLE PHASE UNCERTAINTY - ANOTHER SOURCE OF ERROR IN INDIRECT BLOOD-PRESSURE MEASUREMENT [J].
HATSELL, CP .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1992, 16 (04) :157-158
[3]   RECOMMENDATIONS ON BLOOD-PRESSURE MEASUREMENT [J].
PETRIE, JC ;
OBRIEN, ET ;
LITTLER, WA ;
DESWIET, M .
BRITISH MEDICAL JOURNAL, 1986, 293 (6547) :611-615
[5]  
SLABY A, 1992, J HUM HYPERTENS, V6, P31
[6]  
CUFFLINK NONINVASIVE