LIPOSOME FLOW-INJECTION IMMUNOASSAY - IMPLICATIONS FOR SENSITIVITY, DYNAMIC-RANGE, AND ANTIBODY REGENERATION

被引:94
作者
LOCASCIOBROWN, L
PLANT, AL
HORVATH, V
DURST, RA
机构
[1] National Institute of Standards and Technology, Gaithersburg
[2] Institute for General and Analytical Chemistry, Technical University of Budapest, Budapest Hungary
关键词
D O I
10.1021/ac00222a013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We have developed a liposome-based flow Injection Immunoassay (FI I A) system for quantitation of a clinical analyte, theophylline. With very minor changes In assay format, this procedure can also be used for the quantitation of antl-theophylllne. Automated sequential analyses were performed at room temperature with plcomole sensitivity and a day-to-day coefficient of variation of less than 5% for aqueous solutions. The system components Include liposomes that contain fluorophores In their aqueous centers and an immobllized-antlbody reactor column. The Immunoreactor was regenerated hundreds of times over 3 months of continuous use with no measurable loss of antibody activity. The two assay formats studied produced distinct dynamic ranges for their respective analytes. The special advantages of using flow Injection analysis for Immunoassays and of using liposomes In FI IA are discussed. © 1990, American Chemical Society. All rights reserved.