CELLULOSE SPONGE PUNCH FOR CONTROLLED MITOMYCIN APPLICATION

被引:7
作者
SHIN, DH [1 ]
REED, SY [1 ]
SWORDS, RC [1 ]
REED, AH [1 ]
SIMONE, PA [1 ]
BIRT, CM [1 ]
机构
[1] WAYNE STATE UNIV,SCH MED,DEPT COMMUNITY MED,DETROIT,MI 48201
关键词
D O I
10.1001/archopht.1994.01090240130041
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We describe a new punch for cellulose sponges that helps to standardize the application of mitomycin during glaucoma filtration surgery. The punch was used to create 40 cellulose application disks. The height, diameter, and weight were measured in both wet and dry states. The disk dimensions and weights were reproducible, with a coefficient of variation less than 5 for all categories. The punch has been very useful in our clinical practice and research.
引用
收藏
页码:1624 / 1625
页数:2
相关论文
共 7 条
[1]   TRABECULECTOMY WITH SIMULTANEOUS TOPICAL APPLICATION OF MITOMYCIN-C IN REFRACTORY GLAUCOMA [J].
CHEN, CW ;
HUANG, HT ;
BAIR, JS ;
LEE, CC .
JOURNAL OF OCULAR PHARMACOLOGY, 1990, 6 (03) :175-182
[2]  
CHEN CW, 1986, 25 ACT CONC OPHTH RO, V2, P1487
[3]  
Chen CW, 1983, T ASIA PACIFIC ACAD, V9, P172
[4]   TRABECULECTOMY WITH MITOMYCIN - A COMPARATIVE-STUDY WITH FLUOROURACIL [J].
KITAZAWA, Y ;
KAWASE, K ;
MATSUSHITA, H ;
MINOBE, M .
ARCHIVES OF OPHTHALMOLOGY, 1991, 109 (12) :1693-1698
[5]  
PALMER SS, 1991, OPHTHALMOLOGY, V98, P317
[6]  
SKUTA GL, 1992, OPHTHALMOLOGY, V99, P438
[7]  
WAND M, 1993, OPHTHALMIC SURG LAS, V24, P708