COMPARISON OF MONOPHASIC AND BIPHASIC DEFIBRILLATING PULSE WAVE-FORMS FOR TRANSTHORACIC CARDIOVERSION

被引:78
作者
GREENE, HL
DIMARCO, JP
KUDENCHUK, PJ
SCHEINMAN, MM
TANG, ASL
REITER, MJ
ECHT, DS
CHAPMAN, PD
JAZAYERI, MR
CHAPMAN, FW
AHMED, M
JOHNSON, JL
NISKANEN, RA
机构
[1] UNIV WASHINGTON,SEATTLE,WA 98195
[2] UNIV VIRGINIA,CHARLOTTESVILLE,VA
[3] UNIV CALIF SAN FRANCISCO,SAN FRANCISCO,CA 94143
[4] UNIV OTTAWA,OTTAWA,ON,CANADA
[5] UNIV COLORADO,DENVER,CO 80202
[6] VANDERBILT UNIV,NASHVILLE,TN
[7] ST JOSEPHS HOSP,MILWAUKEE,WI
[8] MT SINAI MED CTR,MILWAUKEE,WI
[9] PHYSIOCONTROL CORP,REDMOND,WA
关键词
D O I
10.1016/S0002-9149(99)80745-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
All transthoracic defibrillators on the US market use nominally monophasic shock waveforms. However, biphasic waveforms have a lower defibrillation threshold than monophasic waveforms for transthoracic defibrillation of animals and for defibrillation of humans by implantable cardioverter defibrillators. The relative efficacies of Edmark monophasic and Gurvich biphasic transthoracic cardioversion waveforms (200 J into 50 Omega) were compared for transthoracic cardioversion in 171 patients undergoing electrophysiologic study for evaluation of ventricular arrhythmias. Patients were randomized in a blinded fashion to receive either a monophasic or a biphasic waveform for the initial shock for conversion of induced ventricular arrhythmias [ventricular fibrillation [VF] = 53, monomorphic ventricular tachycardia [VT] = 80, polymorphic VT = 30, ventricular flutter = 8). Delivered energies for the Edmark and Gurvich waveforms were 215 +/- 11 and 171 +/- 11 J, respectively. There were no significant differences in patient characteristics, use of antiarrhythmic agents, arrhythmia cycle length or duration of arrhythmia prior to shock for monophasic and biphasic waveform groups. The first shock for all arrhythmias was successful in 75 of 88 patients (85.2%) for the monophasic waveform compared with 81 of 83 patients (97.6%) for the biphasic waveform, p = 0.0054. The first shock for VF was successful in 22 of 28 patients (78.6%) for the monophasic waveform compared with 25 of 25 (100%) for the biphasic waveform, p = 0.0241. The Gurvich biphasic waveforms delivering a mean of 171 J were superior to Edmark monophasic waveforms delivering a mean of 215 J for transthoracic cardioversion of arrhythmias of short duration. This finding may have important implications for the development of future transthoracic defibrillators.
引用
收藏
页码:1135 / 1139
页数:5
相关论文
共 18 条
[1]   A PROSPECTIVE RANDOMIZED EVALUATION OF BIPHASIC VERSUS MONOPHASIC WAVEFORM PULSES ON DEFIBRILLATION EFFICACY IN HUMANS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
MEHRA, R ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :728-733
[2]   COMPARISON OF MONOPHASIC WITH SINGLE AND DUAL CAPACITOR BIPHASIC WAVEFORMS FOR NONTHORACOTOMY CANINE INTERNAL DEFIBRILLATION [J].
CHAPMAN, PD ;
VETTER, JW ;
SOUZA, JJ ;
WETHERBEE, JN ;
TROUP, PJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (01) :242-245
[3]   IMPROVED INTERNAL DEFIBRILLATION EFFICACY WITH A BIPHASIC WAVEFORM [J].
FAIN, ES ;
SWEENEY, MB ;
FRANZ, MR .
AMERICAN HEART JOURNAL, 1989, 117 (02) :358-364
[4]   STRENGTH DURATION AND PROBABILITY OF SUCCESS CURVES FOR DEFIBRILLATION WITH BIPHASIC WAVE-FORMS [J].
FEESER, SA ;
TANG, ASL ;
KAVANAGH, KM ;
ROLLINS, DL ;
SMITH, WM ;
WOLF, PD ;
IDEKER, RE .
CIRCULATION, 1990, 82 (06) :2128-2141
[5]   SUPERIORITY OF BIPHASIC SHOCKS IN THE DEFIBRILLATION OF DOGS BY EPICARDIAL PATCHES AND CATHETER ELECTRODES [J].
FLAKER, GC ;
SCHUDER, JC ;
MCDANIEL, WC ;
STOECKLE, H ;
DBEIS, M .
AMERICAN HEART JOURNAL, 1989, 118 (02) :288-291
[6]  
GURVICH N L, 1947, Am Rev Sov Med, V4, P252
[7]   COMPARISON OF THE INTERNAL DEFIBRILLATION THRESHOLDS FOR MONOPHASIC AND DOUBLE AND SINGLE CAPACITOR BIPHASIC WAVEFORMS [J].
KAVANAGH, KM ;
TANG, ASL ;
ROLLINS, DL ;
SMITH, WM ;
IDEKER, RE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (05) :1343-1349
[8]   VENTRICULAR-TACHYCARDIA RATE AND MORPHOLOGY DETERMINE ENERGY AND CURRENT REQUIREMENTS FOR TRANSTHORACIC CARDIOVERSION [J].
KERBER, RE ;
KIENZLE, MG ;
OLSHANSKY, B ;
WALDO, AL ;
WILBER, D ;
CARLSON, MD ;
ASCHOFF, AM ;
BIRGER, S ;
FUGATT, L ;
WALSH, S ;
ROCKWELL, M ;
CHARBONNIER, F .
CIRCULATION, 1992, 85 (01) :158-163
[9]   CRITERIA OF EFFICIENCY AND SAFETY OF THE DEFIBRILLATING IMPULSE [J].
NEGOVSKY, VA ;
SMERDOV, AA ;
TABAK, VY ;
VENIN, IV ;
BOGUSHEVICH, MS .
RESUSCITATION, 1980, 8 (01) :53-67
[10]   TRANS-THORACIC VENTRICULAR DEFIBRILLATION IN THE 100-KG CALF WITH SYMMETRICAL ONE-CYCLE BIDIRECTIONAL RECTANGULAR WAVE STIMULI [J].
SCHUDER, JC ;
GOLD, JH ;
STOECKLE, H ;
MCDANIEL, WC ;
CHEUNG, KN .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (07) :415-422