COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SYSTEM WITH A LASER-INDUCED FLUOROMETRIC DETECTOR FOR DIRECT, AUTOMATED-ASSAY OF SALIVARY CORTISOL

被引:28
作者
OKUMURA, T
NAKAJIMA, Y
TAKAMATSU, T
MATSUOKA, M
机构
[1] Analytical Research Group, Safety and Analytical Research Center, R and D Headquarters, Kouhoku-ku, Yokohama, 223
来源
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS | 1995年 / 670卷 / 01期
关键词
D O I
10.1016/0378-4347(95)00132-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In order to measure human stress, an easy and rapid, fully automated method for the determination of cortisol in saliva has been developed, using column-switching high-performance liquid chromatography with laser-induced fluorescence detection, which involves post-column labeling with sulfuric acid. The developed system requires only 0.1 mi of saliva, and a simple pretreatment consisting of dilution and filtration is sufficient. The column-switching system consisted of a Polymer-Coated Mixed-Functional silica (PCMF) column for deproteinization, and a CN column for frontal concentration and separation. An ODS column in place of the CN provided a better separation, but required a post-column make-up of water for safe reaction. Detection limit of cortisol was 8 fmol (signal-to-noise ratio = 3), which is adequate for routine determination of normal levels of cortisol (1-20 pmol/ml). The analysis time was about 40 min and reproducibility was excellent with an R.S.D. of less than 5%.
引用
收藏
页码:11 / 20
页数:10
相关论文
共 30 条
[1]   EFFECT OF CENTRAL NORADRENALINE DEPLETION ON CORTICOSTERONE LEVELS AND GASTRIC ULCERATIONS IN RATS [J].
BAKKE, HK ;
MURISON, R ;
WALTHER, B .
BRAIN RESEARCH, 1986, 368 (02) :256-261
[2]  
BIDARD M, 1991, ANALUSIS, V19, P302
[3]   DETERMINATION OF SERUM CORTISOL BY THERMOSPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY - COMPARISON WITH GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
GASKELL, SJ ;
ROLLINS, K ;
SMITH, RW ;
PARKER, CE .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1987, 14 (12) :717-722
[4]  
GOTELLI GR, 1981, CLIN CHEM, V27, P441
[5]   ULTRASENSITIVE DIFFERENTIAL MEASUREMENT OF CORTISOL AND CORTISONE IN BIOLOGICAL SAMPLES USING FLUORESCENT ESTER DERIVATIVES IN NORMAL PHASE HPLC [J].
HAEGELE, AD ;
WADE, SE .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1991, 14 (06) :1133-1148
[6]   ASSAY OF HUMAN PLASMA CORTISONE BY LIQUID-CHROMATOGRAPHY - NORMAL PLASMA-CONCENTRATIONS (BETWEEN 8 AND 10 AM) OF CORTISONE AND CORTICOSTERONE [J].
HARIHARAN, M ;
NAGA, S ;
VANNOORD, T ;
KINDT, EK .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1993, 613 (02) :195-201
[7]  
JENKINS GN, 1978, PHYSL BIOCH MOUTH, P213
[8]   SYNTHESIS OF POLYMER-COATED MIXED-FUNCTIONAL PACKING MATERIALS FOR DIRECT ANALYSIS OF DRUG-CONTAINING SERUM AND PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
KANDA, T ;
KUTSUNA, H ;
OHTSU, Y ;
YAMAGUCHI, M .
JOURNAL OF CHROMATOGRAPHY A, 1994, 672 (1-2) :51-57
[9]   DETERMINATION OF CORTICOSTEROIDS IN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AFTER PRECOLUMN DERIVATIZATION WITH 2-(4-CARBOXYPHENYL)-5,6-DIMETHYLBENZIMIDAZOLE [J].
KATAYAMA, M ;
MASUDA, Y ;
TANIGUCHI, H .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1993, 612 (01) :33-39
[10]  
KOBAYASHI Y, 1978, STEROIDS, V32, P137