MONITORING LUNG COMPLIANCE AND END-TIDAL OXYGEN-CONTENT FOR THE DETECTION OF VENOUS AIR-EMBOLISM

被引:12
作者
KYTTA, J
RANDELL, T
TANSKANEN, P
KAJIMOTO, Y
ROSENBERG, PH
机构
[1] HELSINKI UNIV HOSP,DEPT ANAESTHESIOL,HELSINKI,FINLAND
[2] HELSINKI UNIV HOSP,DEPT NEUROSURG,HELSINKI,FINLAND
关键词
MONITORING; OXYGEN; LUNG; COMPLIANCE; SPIROMETRY; EMBOLISM; AIR;
D O I
10.1093/bja/75.4.447
中图分类号
R614 [麻醉学];
学科分类号
100217 [麻醉学];
摘要
Venous air embolism (VAE) is a recognized complication of surgery performed with the patient in the sitting position, but it occurs also during other operations. We report two cases of VAE, associated with a notable decrease in dynamic lung compliance, detected by side-stream spirometry. Based on these cases, an experiment with 10 pigs was designed to evaluate the usefulness of side-stream spirometry in the diagnosis of VAE. Three doses of air (0.5, 1.0 and 2.0 ml kg(-1)) were injected via the proximal part of a 5-French gauge pulmonary artery catheter. Only the largest dose was followed by haemodynamic deterioration. Significant increases in end-tidal oxygen content and decreases in dynamic lung compliance were detected with all doses of air together with conventional signs of VAE, that is increases in pulmonary artery pressures and arterial carbon dioxide tensions, and decreases in end-tidal concentration of carbon dioxide. We conclude that continuous monitoring of end-tidal oxygen concentration and side-stream spirometry offers valuable supplements to other monitoring techniques in the detection of VAE.
引用
收藏
页码:447 / 451
页数:5
相关论文
共 10 条
[1]
CONTINUOUS SPIROMETRY - AN AID TO MONITORING VENTILATION DURING OPERATION [J].
BARDOCZKY, GI ;
ENGELMAN, E ;
DHOLLANDER, A .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 71 (05) :747-751
[2]
DASH HH, 1993, J NEUROSURG ANESTH, V5, P159
[3]
COMPARISON OF VENOUS AIR-EMBOLISM MONITORING METHODS IN SUPINE DOGS [J].
ENGLISH, JB ;
WESTENSKOW, D ;
HODGES, MR ;
STANLEY, TH .
ANESTHESIOLOGY, 1978, 48 (06) :425-429
[4]
TRANSESOPHAGEAL ECHOCARDIOGRAPHY AND TRANS-CUTANEOUS O-2 AND CO2 MONITORING FOR DETECTION OF VENOUS AIR-EMBOLISM [J].
GLENSKI, JA ;
CUCCHIARA, RF ;
MICHENFELDER, JD .
ANESTHESIOLOGY, 1986, 64 (05) :541-545
[5]
INSPIRATORY END-TIDAL OXYGEN-CONTENT DIFFERENCE - A SENSITIVE INDICATOR OF HYPOVENTILATION [J].
LINKO, K ;
PALOHEIMO, M .
CRITICAL CARE MEDICINE, 1989, 17 (04) :345-348
[6]
ANESTHESIA AND SURGERY IN THE SEATED POSITION - ANALYSIS OF 554 CASES [J].
MATJASKO, J ;
PETROZZA, P ;
COHEN, M ;
STEINBERG, P .
NEUROSURGERY, 1985, 17 (05) :695-702
[7]
Schubert A, 1986, ANESTHESIA ANALGESIA, V65, pS135
[8]
DETECTION OF VENOUS AIR-EMBOLISM BY AIRWAY PRESSURE MONITORING [J].
SLOAN, TB ;
KIMOVEC, MA .
ANESTHESIOLOGY, 1986, 64 (05) :645-647
[9]
SPIESS BD, 1986, ANESTHESIA ANALGESIA, V65, pS150
[10]
THE SITTING POSITION IN NEUROSURGERY - A RETROSPECTIVE ANALYSIS OF 488 CASES [J].
STANDEFER, M ;
BAY, JW ;
TRUSSO, R .
NEUROSURGERY, 1984, 14 (06) :649-658