FEEDBACK-CONTROL OF CORONAL PLANE HIP ANGLE IN PARAPLEGIC SUBJECTS USING FUNCTIONAL NEUROMUSCULAR STIMULATION

被引:71
作者
ABBAS, JJ
CHIZECK, HJ
机构
[1] CLEVELAND VET AFFAIRS MED CTR,MOT STUDY LAB,CLEVELAND,OH 44106
[2] CASE WESTERN RESERVE UNIV,DEPT SYST ENGN,CLEVELAND,OH 44106
关键词
D O I
10.1109/10.83570
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports on an investigation of feedback control of coronal plane posture in paraplegic subjects who stand using functional neuromuscular stimulation (FNS). A feedback control system directed at regulating coronal plane hip angle in neutral position was designed, implemented, and evaluated in two paraplegic subjects. The control system included sensor mounting and signal processing techniques, a two-stage feedback controller, stimulation hardware, and a set of percutaneous intramuscular electrodes. The feedback controller consisted of two-stages in cascade: a modified discrete-time proportional-integral-derivative (PID) stage and a nonlinear single-input, multiple-output stage to determine the stimulation to be sent to several muscles. The focus of this work was on evaluating the performance of the feedback controller by comparing the response of the feedback-controlled system to that of an open-loop stimulation system. In an evaluation based on temporal response characteristics the controlled system exhibited a 41% reduction in root-mean-squared (rms) error (where error is defined as the deviation from the desired angle), a 52% reduction in steady-state error,and a 22% reduction in hip compliance. In addition, the feedback-controlled system exhibited significant reductions in variability of these measures on several days. These results demonstrate the ability of the feedback controller to improve the temporal response characteristics of the FNS control system.
引用
收藏
页码:687 / 698
页数:12
相关论文
共 33 条
[1]   A SOFTWARE STRUCTURE FOR IMPLEMENTING MULTI-STATE FEEDBACK CONTROLLERS IN FNS SYSTEMS [J].
ABBAS, JJ ;
CHIZECK, HJ ;
BORGES, G ;
CHOW, P ;
LAMBERT, P ;
MOYNAHAN, M .
PROCEEDINGS OF THE ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, PTS 1-4, 1988, :1653-1654
[2]  
ABBAS JJ, 1987, 40TH P ACEMB NIAG FA, P222
[3]   FNS CONTROL SCHEMES FOR THE UPPER LIMB [J].
ALLIN, J ;
INBAR, GF .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (09) :818-828
[4]  
ANDREWS BJ, 1986, J PHYSL, V380, P19
[5]   CHARACTERIZATION AND CONTROL OF MUSCLE RESPONSE TO ELECTRICAL-STIMULATION [J].
BAJZEK, TJ ;
JAEGER, RJ .
ANNALS OF BIOMEDICAL ENGINEERING, 1987, 15 (05) :485-501
[6]   ADAPTIVE-CONTROL OF ELECTRICALLY STIMULATED MUSCLE [J].
BERNOTAS, LA ;
CRAGO, PE ;
CHIZECK, HJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (02) :140-147
[7]   DEVELOPMENT AND OPERATION OF PORTABLE AND LABORATORY ELECTRICAL-STIMULATION SYSTEMS FOR WALKING IN PARAPLEGIC SUBJECTS [J].
BORGES, G ;
FERGUSON, K ;
KOBETIC, R .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (07) :798-801
[8]   DYNAMIC PERFORMANCE-CHARACTERISTICS OF THE LIQUID-METAL STRAIN-GAUGE [J].
BROWN, TD ;
SIGAL, L ;
NJUS, GO ;
NJUS, NM ;
SINGERMAN, RJ ;
BRAND, RA .
JOURNAL OF BIOMECHANICS, 1986, 19 (02) :165-173
[9]   CONTROL OF FUNCTIONAL NEUROMUSCULAR STIMULATION SYSTEMS FOR STANDING AND LOCOMOTION IN PARAPLEGICS [J].
CHIZECK, HJ ;
KOBETIC, R ;
MARSOLAIS, EB ;
ABBAS, JJ ;
DONNER, IH ;
SIMON, E .
PROCEEDINGS OF THE IEEE, 1988, 76 (09) :1155-1165
[10]  
CHIZECK HJ, 1988, IEEE T BIOMED ENG, V35