DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM

被引:136
作者
HOBAN, PW [1 ]
HEYDARIAN, M [1 ]
BECKHAM, WA [1 ]
BEDDOE, AH [1 ]
机构
[1] UNIV ADELAIDE,DEPT PHYS & MATH PHYS,ADELAIDE,SA 5001,AUSTRALIA
关键词
D O I
10.1088/0031-9155/39/8/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The dose rate dependence and current/voltage characteristics of a PTW Riga diamond detector in the dosimetry of a 6 MV photon beam have been investigated. Diamond detectors are radiosensitive resistors whose conductivity (i) varies almost in proportion to dose rate and (ii) is almost independent of bias voltage for a constant dose rate. At the recommended bias of +100 V, and also at +200 V, the detector is operating with incomplete charge collection due to the electron-hole recombination time being shorter that the maximum time for an electron to be collected by the anode. As dose rate is varied by changing FSD or depth (changing dose per pulse), detector current and dose rate are related by the expression i is-proportional-to D(DELTA) where DELTA is approximately 0.98. This manifests itself in an overestimate in percentage depth-dose at a depth of 30 cm of approximately 1% when compared to ionization chamber results. A similar sublinearity is seen when pulse repetition frequency is varied, indicating that the dependence is an on average rather than an instantaneous dose rate. The dose rate dependence is attributed to the reduction in recombination time as dose rate increases.
引用
收藏
页码:1219 / 1229
页数:11
相关论文
共 11 条
[1]   DOSIMETRY WITH A DIAMOND OPERATING AS A RESISTOR [J].
BURGEMEISTER, EA .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (02) :269-275
[2]   ELECTRICAL-PROPERTIES AND PERFORMANCES OF NATURAL DIAMOND NUCLEAR RADIATION DETECTORS [J].
CANALI, C ;
GATTI, E ;
KOZLOV, SF ;
MANFREDI, PF ;
MANFREDOTTI, C ;
NAVA, F ;
QUIRINI, A .
NUCLEAR INSTRUMENTS & METHODS, 1979, 160 (01) :73-77
[3]  
Fowler JF, 1966, RAD DOSIMETRY
[4]   EVALUATION OF A PTW DIAMOND DETECTOR FOR ELECTRON-BEAM MEASUREMENTS [J].
HEYDARIAN, M ;
HOBAN, PW ;
BECKHAM, WA ;
BORCHARDT, IM ;
BEDDOE, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (08) :1035-1042
[5]  
KHRUNOV VS, 1990, RADIAT PROT DOSIM, V33, P155
[6]  
KONOROVA EA, 1966, FIZ TVERD TELA+, V8, P1
[7]   DIAMOND DOSIMETER FOR X-RAY AND GAMMA-RADIATION [J].
KOZLOV, SF ;
KONOROVA, EA ;
KUZNETSOV, YA ;
SALIKOV, YA ;
REDKO, VI ;
GRINBERG, VR ;
MEILMAN, ML .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (01) :235-237
[8]   PREPARATION AND CHARACTERISTICS OF NATURAL DIAMOND NUCLEAR RADIATION DETECTORS [J].
KOZLOV, SF ;
STUCK, R ;
HAGEALI, M ;
SIFFERT, P .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1975, NS22 (01) :160-170
[9]   EVALUATION OF DIAMOND RADIATION DOSIMETERS [J].
PLANSKOY, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (03) :519-532
[10]   CHARACTERISTICS OF A P-SI DETECTOR IN HIGH-ENERGY ELECTRON FIELDS [J].
RIKNER, G .
ACTA RADIOLOGICA ONCOLOGY, 1985, 24 (01) :71-74