COLLIMATOR OPTIMIZATION FOR LESION DETECTION INCORPORATING PRIOR INFORMATION ABOUT LESION SIZE

被引:29
作者
MOORE, SC
DEVRIES, DJ
NANDRAM, B
KIJEWSKI, MF
MUELLER, SP
机构
[1] HARVARD UNIV,SCH MED,DEPT RADIOL,BOSTON,MA 02115
[2] WORCESTER POLYTECH INST,DEPT BIOMED ENGN,WORCESTER,MA 01609
[3] WORCESTER POLYTECH INST,DEPT MATH SCI,WORCESTER,MA 01609
[4] BRIGHAM & WOMENS HOSP,DEPT RADIOL,BOSTON,MA 02115
[5] UNIV ESSEN GESAMTHSCH KLIN,W-4300 ESSEN 1,GERMANY
关键词
NUCLEAR MEDICINE; COLLIMATOR; ESTIMATION; DETECTION;
D O I
10.1118/1.597466
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A Bayesian estimator has been developed as a paradigm for human observer performance in detecting lesions of unknown size in a uniform noisy background. The Bayesian observer used knowledge of the range of possible lesion sizes as a prior; its predictions agreed well with the results of a six-observer perceptual study. The average human response to changes in collimator resolution, as measured by the detectability index, dA, was tracked by the Bayesian detector's signal-to-noise ratio (SNR) somewhat better than by two other estimation models based, respectively, on lesser and greater degrees of lesion size uncertainty. As the range of possible lesion sizes increased, the Bayesian detector's SNR decreased and the optimal collimator resolution shifted towards better resolution. An analytic approximation for the variance of lesion activity estimates (which included the same prior) was shown to predict the variance of the Bayesian estimator over a wide range of collimator resolution values. Because the bias of the Bayesian estimator was small (1%), the analytic variance estimate permitted a rapid and convenient prediction of the Bayesian detection SNR. This calculation was then used to optimize the geometric parameters of a two-layer tungsten collimator being constructed from crossed grids for a new imaging detector. A Monte Carlo program was first run to estimate all contributions to the radial point-spread function for collimators of differing tungsten contents and spatial resolution values, imaging 140-keV photons emitted from the center of a 15-cm-diameter, water-filled attenuator. The optimal collimator design for detecting lesions with unknown diameters in the range 2.5-7.5 mm yielded a system resolution of ~8.5-mm FWHM, a geometric collimator efficiency of 1.21×10−4, and a single-septum penetration probability of 1%. © 1995, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:703 / 713
页数:11
相关论文
共 29 条
[1]   EFFICIENCY OF DETECTING CHANGES OF DENSITY IN RANDOM DOT PATTERNS [J].
BARLOW, HB .
VISION RESEARCH, 1978, 18 (06) :637-650
[2]  
Barrett H. H., 1986, Proceedings of the SPIE - The International Society for Optical Engineering, V626, P231, DOI 10.1117/12.975398
[3]   OBJECTIVE ASSESSMENT OF IMAGE QUALITY - EFFECTS OF QUANTUM NOISE AND OBJECT VARIABILITY [J].
BARRETT, HH .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1990, 7 (07) :1266-1278
[4]  
BARRETT HH, 1985, P SOC PHOTO-OPT INST, V535, P65, DOI 10.1117/12.947238
[5]   EFFICIENCY OF HUMAN VISUAL SIGNAL DISCRIMINATION [J].
BURGESS, AE ;
WAGNER, RF ;
JENNINGS, RJ ;
BARLOW, HB .
SCIENCE, 1981, 214 (4516) :93-94
[6]  
BURGESS AE, 1982, J APPL PHOTOGR ENG, V8, P76
[7]   DEVELOPMENT AND VALIDATION OF A MONTE-CARLO SIMULATION OF PHOTON TRANSPORT IN AN ANGER CAMERA [J].
DEVRIES, DJ ;
MOORE, SC ;
ZIMMERMAN, RE ;
MUELLER, SP ;
FRIEDLAND, B ;
LANZA, RC .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (04) :430-438
[8]   MAXIMUM-LIKELIHOOD ESTIMATION OF PARAMETERS OF SIGNAL-DETECTION THEORY AND DETERMINATION OF CONFIDENCE INTERVALS - RATING-METHOD DATA [J].
DORFMAN, DD ;
ALF, E .
JOURNAL OF MATHEMATICAL PSYCHOLOGY, 1969, 6 (03) :487-&
[9]  
FAHEY FH, 1992, J NUCL MED, V33, P1859
[10]   HOTELLING TRACE CRITERION AND ITS CORRELATION WITH HUMAN-OBSERVER PERFORMANCE [J].
FIETE, RD ;
BARRETT, HH ;
SMITH, WE ;
MYERS, KJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :945-953