FILM DOSIMETRY FOR RADIOTHERAPY TREATMENT PLANNING VERIFICATION OF A 6 MV TANGENTIAL BREAST IRRADIATION

被引:22
作者
VANBREE, NAM
IDZES, MHM
HUIZENGA, H
MIJNHEER, BJ
机构
[1] DR DANIEL DEN HOED CANC CTR,3075 EA ROTTERDAM,NETHERLANDS
[2] NETHERLANDS CANC INST,AMSTERDAM,NETHERLANDS
关键词
FILM DOSIMETRY; RADIOTHERAPY TREATMENT PLANNING VERIFICATION;
D O I
10.1016/0167-8140(94)90431-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Film dosimetry has been applied to measure relative dose distributions in an anthropomorphic polystyrene breast phantom having cork lungs, simulating a radiation therapy treatment with 6 MV opposing tangential beams. Measured relative dose distributions showed good reproducibility (about 1.5%, 1 SD) and good agreement (<2%) with calculations performed with a three-dimensional treatment planning system. These results demonstrate that film dosimetry is a useful tool for treatment planning verification.
引用
收藏
页码:251 / 255
页数:5
相关论文
共 5 条
[1]   CLINICAL ELECTRON-BEAM DOSIMETRY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO-25 [J].
KHAN, FM ;
DOPPKE, KP ;
HOGSTROM, KR ;
KUTCHER, GJ ;
NATH, R ;
PRASAD, SC ;
PURDY, JA ;
ROZENFELD, M ;
WERNER, BL .
MEDICAL PHYSICS, 1991, 18 (01) :73-109
[2]   GENERATION AND USE OF MEASUREMENT-BASED 3-D DOSE DISTRIBUTIONS FOR 3-D DOSE CALCULATION VERIFICATION [J].
STERN, RL ;
FRAASS, BA ;
GERHARDSSON, A ;
MCSHAN, DL ;
LAM, KL .
MEDICAL PHYSICS, 1992, 19 (01) :165-173
[3]  
VANBATTUM LJ, 1991, INT J RADIAT ONCOL, V18, P69
[4]   3-DIMENSIONAL DOSE DISTRIBUTION OF TANGENTIAL BREAST TREATMENT - A NATIONAL DOSIMETRY INTERCOMPARISON [J].
VANBREE, NAM ;
VANBATTUM, LJ ;
HUIZENGA, H ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1991, 22 (04) :252-260
[5]   FILM DOSIMETRY OF MEGAVOLTAGE PHOTON BEAMS - A PRACTICAL METHOD OF ISODENSITY-TO-ISODOSE CURVE CONVERSION [J].
WILLIAMSON, JF ;
KHAN, FM ;
SHARMA, SC .
MEDICAL PHYSICS, 1981, 8 (01) :94-98