A NUMERICAL AND EXPERIMENTAL INVESTIGATION OF THE FLOW ACCELERATION REGION PROXIMAL TO AN ORIFICE

被引:24
作者
ANAYIOTOS, AS
PERRY, GJ
MYERS, JG
GREEN, DW
FAN, PH
NANDA, NC
机构
[1] UNIV ALABAMA,DIV CARDIOVASC DIS,HEART STN SWS102,BIRMINGHAM,AL 35233
[2] UNIV ALABAMA,DEPT MECH ENGN,BIRMINGHAM,AL 35233
[3] VET ADM MED CTR,CARDIOL SECT,BIRMINGHAM,AL
关键词
ULTRASOUND ECHOCARDIOGRAPHY; COLOR DOPPLER FLOWMAPPING; LASER DOPPLER VELOCIMETRY; NUMERICAL SIMULATION; HEMODYNAMICS; VALVULAR REGURGITATION; PROXIMAL ORIFICE FLOW FIELD;
D O I
10.1016/0301-5629(94)00141-Y
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Attempts to quantify valvular regurgitation have recently been focused on the proximal orifice flow field. A complete description of the proximal orifice flow field is provided in this investigation. A steady state in vitro model accessible by both color Doppler ultrasound (CDU) and laser Doppler velocimetry (LDV) was utilized. Velocities for varying flow rates and orifices were calculated by finite element modeling (FEM), by LDV and by CDU. The steady flow model was composed of circular orifices of 3, 5 and 10 mm diameters at flow rates from 0.7 to 10 L/min. Regurgitant flow rates were calculated from the proximal CDU data by two separate methods. The first approach utilized angle corrected velocities while the second approach utilized only velocities which did not require angle correction (centerline velocities). Both methods correlated well with known flow rates (y = 0.97x -0.09, r = 0.98, SEE = 0.45, p < 0.0001; and y = 1.0x + 0.07, r = 0.99, SEE = 0.27, p < 0.0001, respectively) and were superior to results obtained by assuming a hemispherical geometry as is done in the aliasing technique. The methodology provides a complete analysis of the proximal flow field and involves fewer geometric assumptions than the aliasing approach. This may prove to be an advantage when analyzing in vivo flow fields with complex, uncertain geometry.
引用
收藏
页码:501 / 516
页数:16
相关论文
共 32 条
[1]  
Barclay S A, 1993, J Am Soc Echocardiogr, V6, P433
[2]   A NEW METHOD FOR QUANTITATION OF MITRAL REGURGITATION BASED ON COLOR FLOW DOPPLER IMAGING OF FLOW CONVERGENCE PROXIMAL TO REGURGITANT ORIFICE [J].
BARGIGGIA, GS ;
TRONCONI, L ;
SAHN, DJ ;
RECUSANI, F ;
RAISARO, A ;
DESERVI, S ;
VALDESCRUZ, LM ;
MONTEMARTINI, C .
CIRCULATION, 1991, 84 (04) :1481-1489
[3]   COMPUTER-ANALYSIS OF DOPPLER COLOR FLOW MAPPING IMAGES FOR QUANTITATIVE ASSESSMENT OF INVITRO FLUID JETS [J].
BOLGER, AF ;
EIGLER, NL ;
PFAFF, JM ;
RESSER, KJ ;
MAURER, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (02) :450-457
[4]  
CAPE EG, 1988, J AM COLL CARDIOL, V12, P1354
[5]  
CARO CG, 1978, MECHANICS CIRCULATIO
[6]  
CHEN C, 1991, CIRCULATION S2, V84, P637
[7]  
Durst F., 1976, PRINCIPLES PRACTICE
[8]   COLOR DOPPLER ECHOCARDIOGRAPHIC DETERMINATION OF MITRAL REGURGITANT FLOW FROM THE PROXIMAL VELOCITY PROFILE OF THE FLOW CONVERGENCE REGION [J].
GIESLER, M ;
GROSSMANN, G ;
SCHMIDT, A ;
KOCHS, M ;
LANGHANS, J ;
STAUCH, M ;
HOMBACH, V .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (02) :217-224
[9]   COLOR DOPPLER DETERMINATION OF REGURGITANT FLOW - FROM PROXIMAL ISOVELOCITY SURFACE-AREAS TO PROXIMAL VELOCITY PROFILES - AN INVITRO STUDY [J].
GIESLER, MO ;
STAUCH, M .
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 1992, 9 (01) :51-62
[10]   COLOR DOPPLER ASSESSMENT OF MITRAL REGURGITATION WITH ORTHOGONAL PLANES [J].
HELMCKE, F ;
NANDA, NC ;
HSIUNG, MC ;
SOTO, B ;
ADEY, CK ;
GOYAL, RG ;
GATEWOOD, RP .
CIRCULATION, 1987, 75 (01) :175-183