LIMITATIONS AND LATE COMPLICATIONS OF 3RD-GENERATION AUTOMATIC CARDIOVERTER-DEFIBRILLATORS

被引:106
作者
NUNAIN, SO
ROELKE, M
TROUTON, T
OSSWALD, S
KIM, YH
SOSASUAREZ, G
BROOKS, DR
MCGOVERN, B
GUY, M
TORCHIANA, DF
VLAHAKES, GJ
GARAN, H
RUSKIN, JN
机构
[1] MASSACHUSETTS GEN HOSP,BOSTON,MA 02114
[2] HARVARD UNIV,SCH MED,CARDIAC ARRHYTHMIA SERV,BOSTON,MA
[3] HARVARD UNIV,SCH MED,CARDIAC SURG SERV,BOSTON,MA
关键词
PACING; DEFIBRILLATORS; ARRHYTHMIA; DEFIBRILLATION;
D O I
10.1161/01.CIR.91.8.2204
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background This study examines the limitations and complex management problems associated with the use of tiered-therapy, implantable cardioverter-defibrillators (ICDs). Methods and Results The study group comprises the first 154 patients undergoing implantation of tiered-therapy ICDs at our institution. Pulse generators from three different manufacturers were used. In 39 patients, a complete nonthoracotomy lead system was used. The perioperative mortality was 1.3%. Of these 154 patients, 37% experienced late postoperative problems. Twenty-one patients required system revision within 36.5 months (mean, 8.57+/-11.3) of surgery. Reasons for revision were spurious shocks due to electrode fractures (3) or electrode adapter malfunction (2), inadequate signal from endocardial rate-sensing electrodes (3), superior vena cava or right ventricular coil migration (5), failure to correct tachyarrhythmias due to a postimplant rise in defibrillation threshold (5), or pulse generator failure (3). One of these patients required system removal for infection after revision of an endocardial lead. A further 32 patients received inappropriate shocks for atrial fibrillation with a rapid ventricular response or sinus tachycardia. Two of these patients also received shocks for ventricular tachycardia initiated by antitachycardia pacing triggered by atrial fibrillation. Ventricular pacing for bradycardia was associated with inappropriate shocks due to excessive autogain in 2 patients. Conclusions Despite the major diagnostic and therapeutic advantages of tiered-therapy ICDs, a significant proportion of patients continue to experience hardware-related complications or receive inappropriate shocks.
引用
收藏
页码:2204 / 2213
页数:10
相关论文
共 25 条
[11]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS AND SURVIVAL IN PATIENTS WITH MALIGNANT VENTRICULAR ARRHYTHMIAS [J].
KELLY, PA ;
CANNOM, DS ;
GARAN, H ;
MIRABAL, GS ;
HARTHORNE, JW ;
HURVITZ, RJ ;
VLAHAKES, GJ ;
JACOBS, ML ;
ILVENTO, JP ;
BUCKLEY, MJ ;
RUSKIN, JN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (06) :1278-1286
[12]   EXACERBATION OF VENTRICULAR ARRHYTHMIAS DURING THE POSTOPERATIVE PERIOD AFTER IMPLANTATION OF AN AUTOMATIC DEFIBRILLATOR [J].
KIM, SG ;
FISHER, JD ;
FURMAN, S ;
GROSS, J ;
ZILO, P ;
ROTH, JA ;
FERRICK, KJ ;
BRODMAN, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (05) :1200-1206
[13]   MANAGEMENT OF SURVIVORS OF CARDIAC-ARREST - IS ELECTROPHYSIOLOGIC TESTING OBSOLETE IN THE ERA OF IMPLANTABLE DEFIBRILLATORS [J].
KIM, SG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :756-762
[14]   REDUCTION IN DEFIBRILLATOR SHOCKS WITH AN IMPLANTABLE DEVICE COMBINING ANTITACHYCARDIA PACING AND SHOCK THERAPY [J].
LEITCH, JW ;
GILLIS, AM ;
WYSE, DG ;
YEE, R ;
KLEIN, GJ ;
GUIRAUDON, G ;
SHELDON, RS ;
DUFF, HJ ;
KIESER, TM ;
MITCHELL, LB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (01) :145-151
[15]   DIVERSE MECHANISMS OF UNEXPECTED CARDIAC-ARREST IN ADVANCED HEART-FAILURE [J].
LUU, M ;
STEVENSON, WG ;
STEVENSON, LW ;
BARON, K ;
WALDEN, J .
CIRCULATION, 1989, 80 (06) :1675-1680
[16]   ICD DATA-STORAGE - VALUE IN ARRHYTHMIA MANAGEMENT [J].
MARCHLINSKI, FE ;
GOTTLIEB, CD ;
SARTER, B ;
FINKLE, J ;
HOOK, B ;
CALLANS, D ;
SCHWARTZMAN, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :527-534
[17]   TERMINATION OF MALIGNANT VENTRICULAR ARRHYTHMIAS WITH AN IMPLANTED AUTOMATIC DEFIBRILLATOR IN HUMAN-BEINGS [J].
MIROWSKI, M ;
REID, PR ;
MOWER, MM ;
WATKINS, L ;
GOTT, VL ;
SCHAUBLE, JF ;
LANGER, A ;
HEILMAN, MS ;
KOLENIK, SA ;
FISCHELL, RE ;
WEISFELDT, ML .
NEW ENGLAND JOURNAL OF MEDICINE, 1980, 303 (06) :322-324
[18]  
PACKER M, 1992, CIRCULATION, V85, P50
[19]  
PAUL VE, 1991, EUR J CARDIAC PACING, V1, P21
[20]   EXPERIENCE WITH 3 DIFFERENT 3RD-GENERATION CARDIOVERTER-DEFIBRILLATORS IN PATIENTS WITH CORONARY-ARTERY DISEASE OR CARDIOMYOPATHY [J].
PORTERFIELD, JG ;
PORTERFIELD, LM ;
SMITH, BA ;
BRAY, L .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (03) :301-304