STANDARD FLOW-TIME WAVE-FORMS FOR TESTING OF PEF METERS

被引:25
作者
HANKINSON, JL
CRAPO, RO
机构
[1] UNIV UTAH, SALT LAKE CITY, UT USA
[2] LATTER DAY ST HOSP, DIV PULM, SALT LAKE CITY, UT 84143 USA
关键词
D O I
10.1164/ajrccm.152.2.7633728
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The American Thoracic Society (ATS) has recommended the use of 24 volume-time waveforms for the testing of spirometers. Although these waveforms include values of peak expiratory flow (PEF), they were not originally intended to test PEF meters but, rather, volume parameters for spirometers. In addition, the practice of using ATS volume-time Waveform 24 with varying multiplying factors does not provide the range of flow-time waveform shapes (rise times) needed to evaluate PEF meters. Accordingly, we have developed a set of 26 flow-time waveforms specifically selected to evaluate PEF meters. PEF and other flow parameters (rise time and time to PEF) can be directly measured from these flow-time waveforms. When PEF determined directly from the flow-time curve was compared with PEF determined indirectly from a volume-time curve (ATS-recommended algorithm with an 80 ms time segment), as much as a 10.7% difference between the two methods was observed using a waveform with a fast rise time. In contrast, there was very little difference between the various methods of deriving PEF for waveforms with slower rise times. These 26 Row-time waveforms provide a means of defining PEF for the testing of software algorithms and the testing of PEF meters with computer-driven mechanical pumps.
引用
收藏
页码:696 / 701
页数:6
相关论文
共 9 条
[1]  
Standardization of spirometry - 1987 update, Am. Rev. Respir. Dis., 136, pp. 1285-1298, (1987)
[2]  
Standardization of spirometry, Am. Rev. Respir. Dis., 119, pp. 831-838, (1979)
[3]  
Hankinson J.L., Gardner R.M., Standard waveforms for spirometer testing, Am. Rev. Respir. Dis., 126, pp. 362-364, (1982)
[4]  
Expert Panel Report on Diagnosis and Management of Asthma, (1992)
[5]  
Lanczos C., Applied Analysis, pp. 321-332, (1956)
[6]  
Hankinson J.L., Viola J.O., Ebeling T.R., Petsonk E.L., BTPS correction for a ceramic flow sensor, Chest, 105, pp. 1481-1482, (1994)
[7]  
Hankinson J.L., Viola J.O., Ebeling T.R.
[8]  
Lemen R.J., Gerdes C.B., Wegmann M.J., Perrin K.J., Frequency spectra of flow and volume events for forced vital capacity, J. Appl. Physiol., 53, pp. 977-984, (1982)
[9]  
Signal Processing Toolbox User's Guide, (1988)