FAST IMAGING USING SUBENCODING DATA SETS FROM MULTIPLE DETECTORS

被引:110
作者
RA, JB
RIM, CY
机构
[1] Department of Electrical Engineering, Korea Advanced Institute of Science and Technology, Taejon
[2] Department of Consumer Electronics Research Lab, Gold Star Co., LTD., Seoul
关键词
FAST IMAGING; MULTIPLE DETECTORS; SUBENCODING; ARRAY COIL;
D O I
10.1002/mrm.1910300123
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new fast imaging method using a subencoding data acquisition scheme and a multiple coll receiver system is proposed and demonstrated. In this method, a set of aliased images are produced from receiver coils by using the subencoded data without sacrificing the desired resolution, and resolved to an aliasing-free image by using the distance-dependent sensitivity information of each coil. The reduction rate of data acquisition time is proportional to the number of receiver coils. This method can be easily adapted to conventional imaging methods including fast imaging to further reduce the scan time.
引用
收藏
页码:142 / 145
页数:4
相关论文
共 17 条
[1]  
Ahn C. B., 1986, IEEE T MED IMAGING, V5, P1
[2]   AN ALGORITHM FOR NMR IMAGING RECONSTRUCTION BASED ON MULTIPLE RF RECEIVER COILS [J].
CARLSON, JW .
JOURNAL OF MAGNETIC RESONANCE, 1987, 74 (02) :376-380
[3]   0.3-SECOND FLASH MRI OF THE HUMAN-HEART [J].
FRAHM, J ;
MERBOLDT, KD ;
BRUHN, H ;
GYNGELL, ML ;
HANICKE, W ;
CHIEN, D .
MAGNETIC RESONANCE IN MEDICINE, 1990, 13 (01) :150-157
[4]  
GYNGELL ML, 1986, 5TH ANN M SOC MAGN R, P666
[5]   FLASH IMAGING - RAPID NMR IMAGING USING LOW FLIP-ANGLE PULSES [J].
HAASE, A ;
FRAHM, J ;
MATTHAEI, D ;
HANICKE, W ;
MERBOLDT, KD .
JOURNAL OF MAGNETIC RESONANCE, 1986, 67 (02) :258-266
[6]   RAPID FOURIER-IMAGING USING STEADY-STATE FREE PRECESSION [J].
HAWKES, RC ;
PATZ, S .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (01) :9-23
[7]   FAST MRI DATA ACQUISITION USING MULTIPLE DETECTORS [J].
HUTCHINSON, M ;
RAFF, U .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (01) :87-91
[8]  
Kelton J.R., 1989, P ISMRM, P1172
[9]   A DECOUPLED COIL DETECTOR ARRAY FOR FAST IMAGE ACQUISITION IN MAGNETIC-RESONANCE-IMAGING [J].
KWIAT, D ;
EINAV, S ;
NAVON, G .
MEDICAL PHYSICS, 1991, 18 (02) :251-265
[10]  
LEE SY, 1988, P SOC MAGN RESON, P959