HIGH-QUALITY RECORDING OF BIOELECTRIC EVENTS .1. INTERFERENCE REDUCTION, THEORY AND PRACTICE

被引:348
作者
VANRIJN, ACM [1 ]
PEPER, A [1 ]
GRIMBERGEN, CA [1 ]
机构
[1] UNIV AMSTERDAM,ACAD MED CTR,MED PHYS LAB,1105 AZ AMSTERDAM,NETHERLANDS
关键词
Bioelectric recordings; Driven right leg; Guarding; Interference; Isolation;
D O I
10.1007/BF02441961
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In the first part of the review the various mechanisms that could be the cause of interference in bioelectric recordings are considered. It is demonstrated that the performance of a good amplifier can be seriously degraded in its functioning if the whole measurement situation is not taken into account. Several techniques used to reduce interference, of which guarding and driven right leg circuits are the most important, are analysed. In the second part of the review some examples of the application of the theory in practical situations are presented. The instrumentation amplifier circuit normally used in bioelectric recordings is improved for measurements under difficult circumstances. Another application is a low-cost 64-channel amplifier for multichannel ECG recordings. The third application is a device that can be added to bioelectric measurement systems and will provide a major reduction in interference. © 1990 IFMBE.
引用
收藏
页码:389 / 397
页数:9
相关论文
共 21 条
[1]   SYSTEMIC AND RANDOM VARIATIONS OF ECG ELECTRODE SYSTEM IMPEDANCE [J].
ALMASI, JJ ;
SCHMITT, OH .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1970, 170 (02) :509-&
[2]   MEASUREMENT OF PATIENT BODY CAPACITANCE AND A METHOD OF PATIENT ISOLATION IN MAINS ENVIRONMENTS [J].
FORSTER, IC .
MEDICAL & BIOLOGICAL ENGINEERING, 1974, 12 (05) :730-732
[3]  
GEDDES L. A., 1966, MED BIOL ENG, V4, P439, DOI 10.1007/BF02476166
[4]  
GEDDES LA, 1966, MED RES ENG, V6, P33
[5]  
GEDDES LA, 1972, ELECTRODES MEASUREME, P44
[6]   IMPEDANCE MEASUREMENT OF INDIVIDUAL SKIN SURFACE ELECTRODES [J].
GRIMNES, S .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1983, 21 (06) :750-755
[7]   LOW-POWER, LOW-NOISE INSTRUMENTATION AMPLIFIER FOR PHYSIOLOGICAL SIGNALS [J].
HAMSTRA, GH ;
PEPER, A ;
GRIMBERGEN, CA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1984, 22 (03) :272-274
[8]   60-HZ INTERFERENCE IN ELECTROCARDIOGRAPHY [J].
HUHTA, JC ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (02) :91-101
[9]   EXTRACELLULAR AMPLIFIER WITH BOOTSTRAPPED INPUT STAGE RESULTS IN HIGH COMMON-MODE REJECTION [J].
MCCLELLAN, AD .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1981, 19 (05) :657-658
[10]  
Morrison R., 1977, GROUNDING SHIELDING, P132