ABUTMENT OF HIGH-ENERGY ELECTRON FIELDS

被引:7
作者
HARMS, WB
PURDY, JA
机构
[1] Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1991年 / 20卷 / 04期
关键词
FIELD ABUTMENT; FIELD GAPING; ELECTRON BEAMS;
D O I
10.1016/0360-3016(91)90033-Z
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The problem of positioning multiple electron fields on the skin surface is complicated by the fact that the scattering properties of electrons result in constriction of the higher value isodoses and bulging out of the lower value isodoses. We have studied the dose distributions in the junction region for adjacent electron fields using our 3-dimensional radiation treatment planning system by investigating the effects of field separation for several field size, air gap, and beam energy combinations. Film dosimetry was used to measure specific test cases. Using our 3-dimensional radiation treatment planning system, we are able to display the calculated dose distribution, calculate dose-volume histograms, and compare calculations with the film dosimetry measurements. Results of our study are presented.
引用
收藏
页码:853 / 858
页数:6
相关论文
共 3 条
[1]   ADJACENT FIELDS OF HIGH-ENERGY X-RAYS AND ELECTRONS - FLAT SURFACES [J].
BAGNE, F .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (06) :1186-1191
[2]   CALCULATION OF THE UNCERTAINTY IN THE DOSE DELIVERED DURING RADIATION-THERAPY [J].
GOITEIN, M .
MEDICAL PHYSICS, 1985, 12 (05) :608-612
[3]   A BEAM-EDGE MODIFIER FOR ABUTTING ELECTRON FIELDS [J].
KALEND, AM ;
ZWICKER, RD ;
WU, A ;
STERNICK, ES .
MEDICAL PHYSICS, 1985, 12 (06) :793-798