A DRUG DISSOLUTION MONITOR EMPLOYING MULTIPLE FIBER OPTIC PROBES AND A UV/VISIBLE DIODE-ARRAY SPECTROPHOTOMETER

被引:17
作者
CHEN, CS [1 ]
BROWN, CW [1 ]
机构
[1] UNIV RHODE ISL,DEPT CHEM,KINGSTON,RI 02881
关键词
DRUG DISSOLUTION; FIBER OPTIC PROBES; UV/VISIBLE DIODE ARRAY SPECTROMETER; PRINCIPAL COMPONENT REGRESSION (PCR); FOURIER TRANSFORM;
D O I
10.1023/A:1018975002025
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A traditional dissolution pumping system was recently replaced with a fiber optic interface between the spectrometer and the samples. However, the system was limited to a single sample vessel. In this study, a dissolution testing system with six vessels connected to a diode array spectrometer via six optical fibers was investigated. A bifurcated fiber optic bundle was used to transfer the light from the source to the dissolution vessels and was networked so that spectra of each sample can be measured periodically. A full spectrum calibration method based on Principal Component Regression (PCR) was used to determine the concentrations of active ingredients and to account for interferences due to excipients in tablet formulations. Results on this new fiber optic interface system are compared with those obtained previously with the traditional pumping system. Standard errors of prediction are between 1.5 and 3.2% using cross-validation and between 1.1 and 1.7% for the direct validation of two active ingredients in two different drug formulations.
引用
收藏
页码:979 / 983
页数:5
相关论文
共 15 条
[1]   AN INTRODUCTION TO MULTIVARIATE CALIBRATION AND ANALYSIS [J].
BEEBE, KR ;
KOWALSKI, BR .
ANALYTICAL CHEMISTRY, 1987, 59 (17) :A1007-+
[2]   SEARCHING A UV-VISIBLE SPECTRAL LIBRARY [J].
BROWN, CW ;
DONAHUE, SM .
APPLIED SPECTROSCOPY, 1988, 42 (02) :347-352
[3]   INTERFACING A FIBEROPTIC PROBE TO A DIODE-ARRAY UV-VISIBLE SPECTROPHOTOMETER FOR DRUG DISSOLUTION TESTS [J].
BROWN, CW ;
LIN, J .
APPLIED SPECTROSCOPY, 1993, 47 (05) :615-618
[4]   INFRARED QUANTITATIVE-ANALYSIS IN THE FOURIER DOMAIN - PROCESSING VECTOR REPRESENTATIONS [J].
BROWN, CW ;
OBREMSKI, RJ ;
ANDERSON, P .
APPLIED SPECTROSCOPY, 1986, 40 (06) :734-742
[5]   COMPUTERIZED AUTOMATED-SYSTEM FOR DETERMINING DISSOLUTION RATE PROFILES FOR SOLID DOSAGE FORMS [J].
CIOFFI, FJ ;
ABDOU, HM ;
WARREN, AT .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1976, 65 (08) :1234-1240
[6]   SUCCESSIVE AVERAGE ORTHOGONALIZATION OF SPECTRAL DATA [J].
DONAHUE, SM ;
BROWN, CW .
ANALYTICAL CHEMISTRY, 1991, 63 (10) :980-985
[7]   OPTICAL FIBER SPECTROMETRY IN TURBID SOLUTIONS BY MULTIVARIATE CALIBRATION APPLIED TO TABLET DISSOLUTION TESTING [J].
JOSEFSON, M ;
JOHANSSON, E ;
TORSTENSSON, A .
ANALYTICAL CHEMISTRY, 1988, 60 (24) :2666-2671
[8]  
LAMPARTER E, 1990, GIT FACHZ LAB, V3, P313
[9]   NEAR-IR FIBEROPTIC TEMPERATURE SENSOR [J].
LIN, J ;
BROWN, CW .
APPLIED SPECTROSCOPY, 1993, 47 (01) :62-68
[10]   AUTOMATED DRUG DISSOLUTION MONITOR THAT USES A UV-VISIBLE DIODE-ARRAY SPECTROPHOTOMETER [J].
LO, SC ;
DONAHUE, SM ;
BROWN, CW .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1993, 82 (04) :350-354