THE 200-MEV PROTON THERAPY PROJECT AT THE PAUL-SCHERRER-INSTITUTE - CONCEPTUAL DESIGN AND PRACTICAL REALIZATION

被引:478
作者
PEDRONI, E
BACHER, R
BLATTMANN, H
BOHRINGER, T
CORAY, A
LOMAX, A
LIN, SX
MUNKEL, G
SCHEIB, S
SCHNEIDER, U
TOUROVSKY, A
机构
[1] Department of Radiation Medicine, Paul Scherrer Institute, CH-5232, Villigen PSl
关键词
PROTON THERAPY; DYNAMIC SCANNING; DOSE CONFORMATION; PROTON GANTRY; PRECISION THERAPY;
D O I
10.1118/1.597522
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The new proton therapy facility is being assembled at the Paul Scherrer Institute (PSI). The beam delivered by the PSI sector cyclotron can be split and brought into a new hall where it is degraded from 590 MeV down to an energy in the range of 85–270 MeV. A new beam line following the degrader is used to clean the low-energetic beam in phase space and momentum band. The analyzed beam is then injected into a compact isocentric gantry, where it is applied to the patient using a new dynamic treatment modality, the so-called spot-scanning technique. This technique will permit full three-dimensional conformation of the dose to the target volume to be realized in a routine way without the need for individualized patient hardware like collimators and compensators. By combining the scanning of the focused pencil beam within the beam optics of the gantry and by mounting the patient table eccentrically on the gantry, the diameter of the rotating structure has been reduced to only 4 m. In the article the degrees of freedom available on the gantry to apply the beam to the patient (with two rotations for head treatments) are also discussed. The devices for the positioning of the patient on the gantry (x rays and proton radiography) and outside the treatment room (the patient transporter system and the modified mechanics of the computer tomograph unit) are briefly presented. The status of the facility and first experimental results are introduced for later reference. © 1995, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:37 / 53
页数:17
相关论文
共 24 条
[1]   PROTON RADIATION-THERAPY [J].
ARCHAMBEAU, JO ;
BENNETT, GW ;
LEVINE, GS ;
COWEN, R ;
AKANUMA, A .
RADIOLOGY, 1974, 110 (02) :445-457
[2]   BEAM DELIVERY SYSTEMS FOR CHARGED-PARTICLES [J].
BLATTMANN, H .
RADIATION AND ENVIRONMENTAL BIOPHYSICS, 1992, 31 (03) :219-231
[3]  
BOESECKE R, 1987, USE COMPUTERS RAD TH, P181
[4]   CURRENT DEVELOPMENTS IN PROTON THERAPY - A REVIEW [J].
BONNETT, DE .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (10) :1371-1392
[5]   OPTIMIZATION OF PROTON AND HEAVY-ION THERAPY USING AN ADAPTIVE INVERSION ALGORITHM [J].
BRAHME, A ;
KALLMAN, P ;
LIND, BK .
RADIOTHERAPY AND ONCOLOGY, 1989, 15 (02) :189-197
[6]  
DAUM M, 1989, PSI NUCLEAR PARTICLE, P9
[7]  
EGGER E, 1993, MED RADIOLOGY DIAGNO, P57
[8]  
ESSEN CFV, 1980, J CAN ASSOC RADIOL, V31, P19
[9]  
KOEHLER A M, 1977, Medical Physics (Woodbury), V4, P297, DOI 10.1118/1.594317
[10]   RANGE MODULATORS FOR PROTONS AND HEAVY-IONS [J].
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM .
NUCLEAR INSTRUMENTS & METHODS, 1975, 131 (03) :437-440