A THEORETICALLY AND PRACTICALLY MORE EFFECTIVE METHOD FOR INTERRUPTION OF VENTRICULAR-TACHYCARDIA - SELF-ADAPTING AUTODECREMENTAL OVERDRIVE PACING

被引:65
作者
CHAROS, GS [1 ]
HAFFAJEE, CI [1 ]
GOLD, RL [1 ]
BISHOP, RL [1 ]
BERKOVITS, BV [1 ]
ALPERT, JS [1 ]
机构
[1] UNIV MASSACHUSETTS, MED CTR, DIV CARDIOVASC MED, 55 LAKE AVE N, WORCESTER, MA 01605 USA
关键词
D O I
10.1161/01.CIR.73.2.309
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The efficacy and safety of a new antitachycardia pacing technique, self-adapting decremental overdrive pacing, was assessed in patients with clinical ventricular tachyarrhythmias who underwent programmed ventricular stimulation and serial drug testing. The three phases of this study involved a learning/experience phase, followed by intrapatient comparison of decremental overdrive pacing with conventional antitachycardia pacing modalities of overdrive burst ventricular pacing, and diasotlic scanning with single (S2) and double (S2S3) ventricular extrastimuli. The final phase involved an intrapatient comparison of automated decremental overdrive pacing with overdrive burst ventricular pacing in patients with ventricular tachycardia (VT) cycle lengths of 280 msec or greater. Decremental overdrive pacing was superior to overdrive burst pacing and diastolic scanning (S2S3 and S2) (83% vs 38%, 50%, 9%) in patients with VT cycle lengths of 280 msec or greater. Automated decremental overdrive pacing as applied in the final phase was the most efficacious modality, terminating 92% of VT episodes compared with 56% for overdrive burst pacing in the same patients.
引用
收藏
页码:309 / 315
页数:7
相关论文
共 14 条
[1]  
CICCONE DM, 1985, CIRCULATION, V71, P571
[2]   POTENTIAL HAZARDS OF IMPLANTED DEVICES FOR THE ELECTRICAL CONTROL OF TACHYARRHYTHMIAS [J].
ECHT, DS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03) :580-587
[3]   TERMINATION OF VENTRICULAR TACHYCARDIA WITH BURSTS OF RAPID VENTRICULAR PACING [J].
FISHER, JD ;
MEHRA, R ;
FURMAN, S .
AMERICAN JOURNAL OF CARDIOLOGY, 1978, 41 (01) :94-102
[4]   CARDIAC PACING AND PACEMAKERS .2. SERIAL ELECTROPHYSIOLOGIC-PHARMACOLOGIC TESTING FOR CONTROL OF RECURRENT TACHYARRHYTHMIAS [J].
FISHER, JD ;
COHEN, HL ;
MEHRA, R ;
ALTSCHULER, H ;
ESCHER, DJW ;
FURMAN, S .
AMERICAN HEART JOURNAL, 1977, 93 (05) :658-668
[5]   ELECTRICAL TERMINATION OF TACHYARRHYTHMIAS BY DISCRETE PULSES [J].
GERMAN, LD ;
STRAUSS, HC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03) :514-521
[6]   ELECTROPHYSIOLOGIC APPROACH TO THERAPY OF RECURRENT SUSTAINED VENTRICULAR TACHYCARDIA [J].
JOSEPHSON, ME ;
HOROWITZ, LN .
AMERICAN JOURNAL OF CARDIOLOGY, 1979, 43 (03) :631-642
[7]   TRANSIENT ENTRAINMENT AND INTERRUPTION OF VENTRICULAR-TACHYCARDIA [J].
MACLEAN, WAH ;
PLUMB, VJ ;
WALDO, AL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (04) :358-366
[8]   RECURRENT RECIPROCATING TACHYCARDIAS IN WOLFF-PARKINSON-WHITE SYNDROME - CONTROL BY USE OF A SCANNING PACEMAKER [J].
MANDEL, WJ ;
LAKS, MM ;
YAMAGUCHI, I ;
FIELDS, J ;
BERKOVITS, B .
CHEST, 1976, 69 (06) :769-774
[9]   ELECTRODE-CATHETER ARRHYTHMIA INDUCTION IN SELECTION AND ASSESSMENT OF ANTIARRHYTHMIC DRUG-THERAPY FOR RECURRENT VENTRICULAR TACHYCARDIA [J].
MASON, JW ;
WINKLE, RA .
CIRCULATION, 1978, 58 (06) :971-985
[10]   INFLUENCE OF TACHYCARDIA CYCLE LENGTH AND ANTI-ARRHYTHMIC DRUGS ON PACING TERMINATION AND ACCELERATION OF VENTRICULAR-TACHYCARDIA [J].
NACCARELLI, GV ;
ZIPES, DP ;
RAHILLY, GT ;
HEGER, JJ ;
PRYSTOWSKY, EN .
AMERICAN HEART JOURNAL, 1983, 105 (01) :1-5