RAPID IMMUNOFILTRATION ASSAY OF NEWCASTLE-DISEASE VIRUS USING A SILICON SENSOR

被引:19
作者
LEE, WE [1 ]
THOMPSON, HG [1 ]
HALL, JG [1 ]
FULTON, RE [1 ]
WONG, JP [1 ]
机构
[1] UNIV VICTORIA,DEPT BIOCHEM,VICTORIA V8W 3P6,BC,CANADA
关键词
NEWCASTLE DISEASE VIRUS; STREPTAVIDIN; BIOTIN; FILTRATION; DETECTION; BIOSENSOR; LIGHT ADDRESSABLE POTENTIOMETRIC SENSOR; IMMUNOASSAY;
D O I
10.1016/0022-1759(93)90336-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid nonradioactive sandwich immunoassay which utilizes biotin-streptavidin mediated filtration capture of immune complexes in conjunction with a silicon sensor was developed for the detection of virus. Using purified Newcastle disease virus as a model, the lower limits of detection (LOD) were determined for a number of immunoassay configurations employing both monoclonal and polyclonal antibodies. The LODs ranged from 1.3 ng/ml (sample volume of 100 mu l) for an incubation of 60 min to 400 ng/ml for a 1 min incubation. The sandwich immune complexes were formed from one-step incubation of antibody and antigen. No 'hook' effects were observed over a wide range of analyte concentrations. The assays were easy to perform and required a total time equal to the incubation period plus about 5 min. The assay format is suitable for virus, bacteria and protein antigens. New assays can be developed and optimized readily, often within 1 day.