IMPROVED QUANTIFICATION OF RADIONUCLIDE UPTAKE USING DECONVOLUTION AND WINDOWED SUBTRACTION TECHNIQUES FOR SCATTER COMPENSATION IN SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY

被引:33
作者
YANCH, JC
FLOWER, MA
WEBB, S
机构
[1] ROYAL MARSDEN HOSP,SUTTON SM2 5PT,SURREY,ENGLAND
[2] INST CANC RES,JOINT DEPT PHYS,SUTTON SM2 5PX,SURREY,ENGLAND
关键词
D O I
10.1118/1.596453
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A comparison of two methods of scatter compensation in single photon emission computed tomography (SPECT) imaging is made on the basis of improvements in quantification. The methods, scatter-window subtraction and constrained deconvolution of an average point source response function (PSRF), are described; the theoretical basis of each method is also briefly assessed. Improvements in relative quantification offered by each method are measured by examining ratios of counts in hot cylinders to counts in a slightly radioactive background. The cylinder/background concentration ratio was varied by a factor of five; the sizes of the cylinders remained constant. Keeping the diameter of the cylinders constant allowed for an assessment of the effect of concentration differences, without the conflicting effect of variation in the size of the hot source. Results showed that while both scatter-window subtraction and constrained deconvolution offer quantification improvements in the final image, the method of prereconstruction deconvolution of a planar PSRF from each planar projection is substantially more successful than either scatter-window subtraction or other methods of implementing the deconvolution procedure. © 1990, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:1011 / 1022
页数:12
相关论文
共 32 条
[1]  
AXELSSON B, 1984, J NUCL MED, V25, P490
[2]  
BARRETT HH, 1981, RADIOLOGICAL IMAGING, P33
[3]  
BARRETT HH, 1981, RADIOLOGICAL IMAGING, P180
[4]   ASPECTS OF IMAGING AND COUNTING IN NUCLEAR-MEDICINE USING SCINTILLATION AND SEMICONDUCTOR DETECTORS [J].
BECK, RN ;
CHARLESTON, DB ;
ZIMMER, LT ;
HOFFER, PB .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1972, NS19 (03) :173-+
[5]  
BECK RN, 1969, MED RADIOISOTOPE SCI
[6]  
BLOCH P, 1972, J NUCL MED, V14, P67
[7]   CONSTRAINED OPTIMIZATION AND ITS APPLICATION TO SCINTIGRAPHY [J].
BOARDMAN, AK .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (02) :363-371
[8]   STUDY OF THE APPLICATION OF A DECONVOLUTION METHOD TO SCINTIGRAPHY [J].
CORDIER, S ;
BIRAUD, Y ;
CHAMPAILLER, A ;
VOUTAY, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :577-582
[9]  
EHRHARDT JC, 1974, J NUCL MED, V15, P943
[10]   MONTE-CARLO EVALUATION OF COMPTON SCATTER SUBTRACTION IN SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY [J].
FLOYD, CE ;
JASZCZAK, RJ ;
HARRIS, CC ;
GREER, KL ;
COLEMAN, RE .
MEDICAL PHYSICS, 1985, 12 (06) :776-778