An analysis is presented, with supporting experimental measurements, of the non-linear effects in electron intensity measurements using channel electron multipliers (CEMs) in both the analogue amplification and pulse counting modes. It is shown that for Auger electron spectroscopy, where the CEMs may be used in the analogue mode, the output current should not exceed 0.8r% of the CEM wall current for the linearity to be maintained within r%. In the pulse counting mode, used more generally, simple dead time corrections are possible over a limited range of counting rates. This range is defined theoretically and experimental procedures are demonstrated to show how the range may be defined, and the necessary corrections made, to retain accurate linearity within 0.5%.
引用
收藏
页码:240 / 246
页数:7
相关论文
共 10 条
[1]  
BOUTOT JP, 1976, ADV ELECTRONICS EL A, V40, P103
[2]  
Guest A. J., 1971, ACTA ELECTRON, V14, P79
[3]  
HENKEL P, DESIGN PRINCIPLES CH
[4]  
HERMANN KH, 1987, J PHYS E, V20, P177
[5]   STATISTICS OF ELECTRON MULTIPLICATION [J].
LOMBARD, FJ ;
MARTIN, F .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1961, 32 (02) :200-&
[6]   DEAD-TIME PROBLEMS [J].
MULLER, JW .
NUCLEAR INSTRUMENTS & METHODS, 1973, 112 (1-2) :47-57
[7]   CHANNEL ELECTRON MULTIPLIERS - QUANTITATIVE INTENSITY MEASUREMENT - EFFICIENCY, GAIN, LINEARITY AND BIAS EFFECTS [J].
SEAH, MP .
JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA, 1990, 50 (1-2) :137-157
[8]  
1980, 4000 GAL EL CORP CHA
[9]  
1985, CDEM DATA SHEET
[10]  
1987, PHILIPS COMPONENTS 3