THE MICRO WATER 1,2-DICHLOROETHANE INTERFACE AS A TRANSDUCER FOR CREATININE ASSAY

被引:38
作者
OSBORNE, MD
GIRAULT, HH
机构
[1] Institut de Chimie Physique III, Ecole Polytechnique Fédérale de Lausanne, Lausanne
关键词
BIOSENSOR; MICRO-ITIES; CREATININE; AMPEROMETRIC-ISE;
D O I
10.1007/BF01244888
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Facilitated ammonium ion transfer reactions, using the crown ether dibenzo-18-crown-6, at a single microhole supported water/1,2-dichloroethane interface were used as the transducer to detect the biomolecule creatinine. The ammonium ion was produced during the enzymatic action of creatinine deiminase on creatinine. A prototype sensing device incorporating a Teflon gas permeable membrane, to exclude possible interferents, yielded a linear amperometric response to added creatinine from 20 muM to 1 mM. This range is sufficient for that required for the determination of creatinine in serum.
引用
收藏
页码:175 / 185
页数:11
相关论文
共 19 条
[1]   CHARACTERIZATION OF BIOSENSORS BASED ON MEMBRANES CONTAINING A CONDUCTING POLYMER [J].
BATTILOTTI, M ;
COLAPICCHIONI, C ;
GIANNINI, I ;
PORCELLI, F ;
CAMPANELLA, L ;
CORDATORE, M ;
MAZZEI, F ;
TOMASSETTI, M .
ANALYTICA CHIMICA ACTA, 1989, 221 (01) :157-161
[2]   DETERMINATION OF BUTYRYLCHOLINESTERASE INHIBITION USING ION TRANSFER ACROSS THE INTERFACE BETWEEN 2 IMMISCIBLE LIQUIDS [J].
BEATTIE, PD ;
INFELTA, PP ;
GIRAULT, HH .
ANALYTICAL CHEMISTRY, 1994, 66 (01) :52-57
[3]  
GUILBAULT GG, 1977, ANAL CHEM, V49, P795
[4]  
HOWANITZ JH, 1991, LABORATORY MED TEST, P43
[5]  
JESPESSEN MT, 1988, ANAL CHIM ACTA, V214, P147
[6]  
LEHNINGER AL, 1991, BIOCHEMISTRY
[7]   AMMONIA ABATEMENT IN AN ENZYMATIC FLOW SYSTEM FOR THE DETERMINATION OF CREATININE IN BLOOD SERA AND URINE [J].
MASCINI, M ;
FORTUNATI, S ;
MOSCONE, D ;
PALLESCHI, G .
ANALYTICA CHIMICA ACTA, 1985, 171 (MAY) :175-184
[8]   DETERMINATION OF CREATININE IN CLINICAL-SAMPLES WITH A CREATININASE REACTOR AND AN AMMONIA PROBE [J].
MASCINI, M ;
PALLESCHI, G .
ANALYTICA CHIMICA ACTA, 1982, 136 (APR) :69-76
[9]   FLOW-INJECTION POTENTIOMETRIC DETERMINATION OF CREATININE IN URINE USING SUB-NERNSTIAN LINEAR RESPONSE RANGE [J].
MATUSZEWSKI, W ;
TROJANOWICZ, M ;
MEYERHOFF, ME ;
MOSZCZYNSKA, A ;
LANGEMOROZ, E .
ELECTROANALYSIS, 1993, 5 (02) :113-120
[10]  
NGUYEN VK, 1991, ANAL CHEM, V63, P611