COMPARISON OF 7-FRENCH AND 8-FRENCH GUIDING CATHETERS FOR ELECTIVE PTCA - RESULTS OF A PROSPECTIVE RANDOMIZED TRIAL

被引:13
作者
TALLEY, JD
WILKINS, C
CICCONE, J
HATTEL, LJ
机构
[1] UNIV LOUISVILLE,SCH MED,DIV CARDIOVASC,INVAS & INTERVENT CARDIOL SECT,LOUISVILLE,KY 40292
[2] SETON HOSP,AUSTIN,TX
[3] MEM HOSP,CHEYENNE,WY
[4] BETH ISRAEL HOSP,NEWARK,NJ
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1994年 / 33卷 / 03期
关键词
GUIDING CATHETERS; ANGIOPLASTY; PTCA;
D O I
10.1002/ccd.1810330303
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A prospective randomized trial was performed to detect technical and clinical differences of 7F compared to 85 guiding catheters (GC) used in elective coronary angioplasty (PTCA). One hundred and fifteen patients undergoing elective PICA with standard balloon dilatation catheters were randomized to 7F (55 pts) or 8F (58 pts) guiding catheters. The endpoints were primarily coronary artery and peripheral vascular complications; and secondarily, technical details and quantitative and qualitative angiographic quality. There was no difference between 7F or 8F CC for development of coronary or peripheral vascular complications. With 7F GC there was less blood loss (Hct, 3.5 +/- 3.4% vs. 6.5 +/- 9.6%, P = .033), and less contrast medium (160 +/- 88 mt vs. 200 +/- 119 mt, P = .049) used. Angiographic quality was similar, although visualization of lesions in the left anterior descending coronary artery in the left anterior oblique projection was improved with the 8F GC. There is no advantage of 7F GC for the prevention of coronary or peripheral vascular complications, although there was less blood loss and contrast medium used with the 7F systems. (C) 1994 Wiley-Liss, Inc.
引用
收藏
页码:205 / 211
页数:7
相关论文
共 18 条
[1]   EARLY HOSPITAL DISCHARGE AFTER PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
CRAGG, DR ;
FRIEDMAN, HZ ;
ALMANY, SL ;
GANGADHARAN, V ;
RAMOS, RG ;
LEVINE, AB ;
LEBEAU, TA ;
ONEILL, WW .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (19) :1270-1274
[2]  
ELLIS SG, 1994, TXB INTERVENTIONAL C, P186
[3]   CORONARY ANGIOPLASTY USING NEW 6 FRENCH GUIDING CATHETERS [J].
FELDMAN, R ;
GLEMSER, E ;
KAIZER, J ;
STANDLEY, M .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1991, 23 (02) :93-99
[4]  
HASNIE A, 1994, CLIN RES, V42, pA291
[5]  
JOSEPH SA, 1993, J INTERV CARDIOL, V6, P213
[6]   PRELIMINARY EXPERIENCE WITH 5 AND 6 FRENCH DIAGNOSTIC CATHETERS AS GUIDING CATHETERS FOR CORONARY ANGIOPLASTY [J].
KERN, MJ ;
TALLEY, JD ;
DELIGONUL, U ;
SEROTA, H ;
AGUIRRE, F ;
GUDIPATI, C ;
RING, M ;
JOSEPH, A ;
YUSSMAN, ZA ;
KULICK, D ;
SALINGER, M .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1991, 22 (01) :60-63
[7]   CORONARY ANGIOPLASTY THROUGH 4-FRENCH DIAGNOSTIC CATHETERS [J].
MEHAN, VK ;
MEIER, B ;
URBAN, P ;
VERINE, V ;
HAINE, E ;
DORSAZ, PA .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1993, 30 (01) :22-26
[8]   PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY THROUGH 4 FRENCH DIAGNOSTIC CATHETERS [J].
MOLES, VP ;
MEIER, B ;
URBAN, P ;
DELASERNA, F ;
PANDE, AK .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1992, 25 (02) :98-100
[9]   CORONARY ANGIOPLASTY AT THE TIME OF INITIAL CATHETERIZATION USING SMALL DIAGNOSTIC CATHETERS [J].
PANAYIOTOU, H ;
NORRIS, JW ;
FORMAN, MB .
AMERICAN HEART JOURNAL, 1990, 119 (01) :204-205
[10]   PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY THROUGH 6F DIAGNOSTIC CATHETERS - A FEASIBILITY STUDY [J].
RESAR, JR ;
PREWITT, KC ;
WOLFF, MR ;
BLUMENTHAL, R ;
RAQUENO, JV ;
BRINKER, JA .
AMERICAN HEART JOURNAL, 1993, 125 (06) :1591-1596