NEURAL TISSUE COMPATIBILITY OF TEFLON AS AN IMPLANT MATERIAL FOR MICROVASCULAR DECOMPRESSION

被引:23
作者
AMMAR, A
LAGENAUR, C
JANNETTA, P
机构
[1] Department of Neurosurgery, College of Medicine and Medical Sciences, King Faisal University, Dammam
[2] Department of Neurobiology, Anatomy, and Cell Science, University of Pittsburgh, School of Medicine, Pittsburgh, Pennsylvania
[3] Department of Neurological Surgery, University of Pittsburgh, School of Medicine, Pittsburgh, Pennsylvania
关键词
Cell adhesion; cell culture; microvascular decompression; mouse cerebellum; Teflon; tissue compatibility;
D O I
10.1007/BF00346368
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Teflon is utilized in neurosurgery as well as in plastic, vascular and heart surgery. Although the effect of Teflon on different types of cells and tissues has been previously studied, we are not aware of any study in which the effect of Teflon was tested on cells of the central nervous system. We have therefore examined the tissue compatibility of spongy and fibrous Teflon by directly exposing the Teflon to dissociated cerebellar cells containing both glia and neurons in tissue culture. Daily examination of the growth of the cells adjacent to Teflon fibers using an inverted phase contrast microscope revealed that Teflon has little or no effect on the growth of these cells. When the cells are fixed after 7 days in culture and stained by the Jenner-Giemsa method, adhesion of both glia and neurons to the surface of the Teflon was seen. Attachment of neural cells to the Teflon was not extensive, as was shown by indirect immunofluorescence technique in connection with double-label staining with anti-GFAP as glia marker and anti-M6 as mouse neuron marker. Thus, these experiments show that Teflon is relatively inert when used as an implant in the central nervous system. © 1990 Walter de Gruyter & Co.
引用
收藏
页码:299 / 303
页数:5
相关论文
共 29 条
[1]  
Ammar A., Kobayashi S., Sugita K., Shimizu Y., Tissue compatibility of ceramics as compared with other implant materials by means of tissue culture, Proceedings of the First Orthopedic Ceramics Implant Meetings, pp. 25-30, (1981)
[2]  
Ammar A., Tissue compatibility of different intracranial implant materials: in vivo and in-vitro studies, Acta Neurochirur, 72, pp. 45-59, (1984)
[3]  
Bignami A., Eng L.F., Dahl D., Uyeda C.T., Localization of the glial fibrillary acidic protein in astrocytes by immunofluorescence, Brain Res, 43, pp. 429-435, (1972)
[4]  
Boone C.W., Jacobs J.B., Sarcomas routinely produced from putatively nontumorigenic cells by subcutaneous inoculation attached to plastic platelets, J Supramol Struct, 5, pp. 131-137, (1976)
[5]  
Boyd L.K., Schmidt S., Pippert T.R., Hite S.A., Sharp W.V., The effect of pore size and endothelial cell seeding upon the performance of small diameter e-PTFF vascular grafts under controlled flow conditions, J Biomed Mater Res, 22, pp. 163-177, (1988)
[6]  
Cavallini G., Lanfredi M., Lodi M., Govoni M., Pampolini M., Detection and measurement of a cellular immune-reactivity towards polyester and polytetrafluoroethylene grafts, Acta Chir Scand, 153, pp. 179-184, (1987)
[7]  
Cuadros C.L., Granatir C.E., Nerve regeneration through a (synthetic microporous tube expanded) polytetrafluoroethylene: experimental study in the sciatic nerve of the rat, Microsurgery, 8, pp. 41-47, (1987)
[8]  
Eng L.F., Vanderhaeghen J.J., Bignami A., Gerstl B., An acidic protein isolated from fibrous astrocytes, Brain Res, 28, pp. 351-354, (1971)
[9]  
Ericksson A.S., Bjursten L.M., Ericson L.E., Thoen P., Hollow implants in soft tissue allowing quantitative studies of cells and fluid at the implant interface, Biomaterials, 9, pp. 86-90, (1988)
[10]  
Feldman M.D., Sataloff R.T., Chio H.Y., Ballas S.K., Compatibility of autologous fibrin adhesive with implant materials, Arch Otolaryngol Head Neck Surg, 114, pp. 182-185, (1988)