EFFECTIVE DEFIBRILLATION IN PIGS USING INTERLEAVED AND COMMON PHASE SEQUENTIAL BIPHASIC SHOCKS

被引:11
作者
GUSE, PA
ROLLINS, DL
KRASSOWSKA, W
WOLF, PD
SMITH, WM
IDEKER, RE
机构
[1] DUKE UNIV, MED CTR, DEPT MED, POB 3140, DURHAM, NC 27710 USA
[2] DUKE UNIV, MED CTR, DEPT PATHOL, DURHAM, NC 27710 USA
[3] DUKE UNIV, SCH SCI, EMERGING CARDIOVASC TECHNOL ENGN RES CTR, DURHAM, NC 27706 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1993年 / 16卷 / 08期
关键词
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; TRANSVENOUS DEFIBRILLATION; SEQUENTIAL SHOCKS; BIPHASIC SHOCKS;
D O I
10.1111/j.1540-8159.1993.tb01043.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Previous studies have shown that low internal defibrillation thresholds (DFTs) can be attained by using two pairs of electrodes and combining biphasic shocks with sequential timing. The purpose of this two-part study was to test the defibrillation efficacy of two new shock sequences, an interleaved biphasic, and a common phase sequential biphasic, that utilized two pairs of electrodes and were developed from the concept of sequential biphasic shocks. In the first part, defibrillation catheters were placed in the right ventricle and the superior vena cava of six anesthetized pigs. A small patch electrode was placed on the LV apex through a subxiphoid incision and a cutaneous patch was placed on the left thorax. The mean DFT energies for the interleaved biphasic (5.2 +/- 0.4 J) and the common phase sequential biphasic waveforms (5.4 +/- 0.4 J) were substantially less (P < 0.0001) than those for either the sequential monophasic (10.6 +/- 1.0 J) or single biphasic waveforms (9.0 +/- 1.0 J). In the second study, which used nine anesthetized pigs, the importance of phase reversal was demonstrated by the finding that the DFT energy of a common phase sequential biphasic shock (6.2 +/- 0.4 J) was much less than a common phase sequential monophasic shock (17.9 +/- 1.3 J, P < 0.000 1); furthermore, the average DFT for four common phase sequential biphasic configurations (5.7 +/- 0.2 1) was much less than for a configuration that was similar except that current flow was not reversed in one phase so that no biphasic effect was present (19.7 +/- 1.2 J). The efficacy of common phase sequential biphasics was comparable to that of sequential biphasics. The effectiveness of sequential biphasics, interleaved biphasics, and common phase sequential biphasics is possibly due to two mechanisms: (A) an increase in the potential gradient during a later phase in regions that were low during the first phase, and (B) the exposure of most of the myocardium to a biphasic shock that reduces the minimum extracellular potential gradient needed to defibrillate.
引用
收藏
页码:1719 / 1734
页数:16
相关论文
共 47 条
[1]   DIRECTIONAL VARIABILITY OF STIMULATION THRESHOLD MEASUREMENTS IN ISOLATED GUINEA-PIG CARDIOMYOCYTES - RELATIONSHIP WITH ORTHOGONAL SEQUENTIAL DEFIBRILLATING PULSES [J].
BARDOU, AL ;
CHESNAIS, JM ;
BIRKUI, PJ ;
GOVAERE, MC ;
AUGER, PM ;
VONEUW, D ;
DEGONDE, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12) :1590-1595
[2]   A PROSPECTIVE RANDOMIZED EVALUATION OF BIPHASIC VERSUS MONOPHASIC WAVEFORM PULSES ON DEFIBRILLATION EFFICACY IN HUMANS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
MEHRA, R ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :728-733
[3]   AN EFFECTIVE AND ADAPTABLE TRANSVENOUS DEFIBRILLATION SYSTEM USING THE CORONARY SINUS IN HUMANS [J].
BARDY, GH ;
ALLEN, MD ;
MEHRA, R ;
JOHNSON, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (04) :887-895
[4]  
BOURLAND JD, 1986, MED INSTRUM, V20, P138
[5]   COMPARISON OF MONOPHASIC WITH SINGLE AND DUAL CAPACITOR BIPHASIC WAVEFORMS FOR NONTHORACOTOMY CANINE INTERNAL DEFIBRILLATION [J].
CHAPMAN, PD ;
VETTER, JW ;
SOUZA, JJ ;
WETHERBEE, JN ;
TROUP, PJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (01) :242-245
[6]   THE POTENTIAL GRADIENT FIELD CREATED BY EPICARDIAL DEFIBRILLATION ELECTRODES IN DOGS [J].
CHEN, PS ;
WOLF, PD ;
CLAYDON, FJ ;
DIXON, EG ;
VIDAILLET, HJ ;
DANIELEY, ND ;
PILKINGTON, TC ;
IDEKER, RE .
CIRCULATION, 1986, 74 (03) :626-636
[7]   THE EFFECT OF PHASE-SEPARATION ON BIPHASIC WAVE-FORM DEFIBRILLATION [J].
COOPER, RAS ;
WALLENIUS, ST ;
SMITH, WM ;
IDEKER, RE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :471-482
[8]   MYOCARDIAL NECROSIS FROM DIRECT-CURRENT COUNTERSHOCK - EFFECT OF PADDLE ELECTRODE SIZE AND TIME INTERVAL BETWEEN DISCHARGES [J].
DAHL, CF ;
EWY, GA ;
WARNER, ED ;
THOMAS, ED .
CIRCULATION, 1974, 50 (05) :956-961
[9]   THE RELATIONSHIP BETWEEN SUCCESSFUL DEFIBRILLATION AND DELIVERED ENERGY IN OPEN-CHEST DOGS - REAPPRAISAL OF THE DEFIBRILLATION THRESHOLD CONCEPT [J].
DAVY, JM ;
FAIN, ES ;
DORIAN, P ;
WINKLE, RA .
AMERICAN HEART JOURNAL, 1987, 113 (01) :77-84
[10]   IMPROVED DEFIBRILLATION THRESHOLDS WITH LARGE CONTOURED EPICARDIAL ELECTRODES AND BIPHASIC WAVE-FORMS [J].
DIXON, EG ;
TANG, ASL ;
WOLF, PD ;
MEADOR, JT ;
FINE, MJ ;
CALFEE, RV ;
IDEKER, RE .
CIRCULATION, 1987, 76 (05) :1176-1184