OPTICAL OXYGEN SENSOR BASED ON PHOSPHORESCENCE LIFETIME QUENCHING AND EMPLOYING A POLYMER IMMOBILIZED METALLOPORPHYRIN PROBE .1. THEORY AND INSTRUMENTATION

被引:51
作者
GEWEHR, PM
DELPY, DT
机构
[1] UCL, DEPT MED PHYS & BIOENGN, SHROPSHIRE HOUSE, CAPPER ST, LONDON WC1E 6JA, ENGLAND
[2] UNIV LONDON UNIV COLL, CTR MED TECHNOL, LONDON WC1E 6BT, ENGLAND
关键词
LIFETIME; METALLOPORPHYRIN; OPTICAL; OXYGEN; PHOSPHORESCENCE; SENSOR;
D O I
10.1007/BF02446879
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A review of the theory of phosphorescence quenching is given, and its particular application to the sensing of oxygen is outlined. The advantages of measuring phosphorescence lifetime as opposed to phosphorescence intensity are reviewed. The advantages of using the metalloporphyrins as such sensors are identified and in particular the characteristics of palladium coproporphyrin are discussed. The exceptionally long room temperature lifetime of this material makes it possible to use relatively simple PC-based instrumentation to measure lifetimes, with a xenon flash-lamp light source. The design of such a system is given and its performance in measuring phosphorescence lifetime in aqueous solutions is demonstrated