Resolution of semiconductor multilayers using backscattered electrons in scanning electron microscopy

被引:3
作者
Govoni, D [1 ]
Merli, PG [1 ]
Migliori, A [1 ]
Nacucchi, M [1 ]
机构
[1] PASTIS CNRSM SCPA,I-72100 BRINDISI,ITALY
来源
MICROSCOPY MICROANALYSIS MICROSTRUCTURES | 1995年 / 6卷 / 5-6期
关键词
D O I
10.1051/mmm:1995141
中图分类号
TH742 [显微镜];
学科分类号
摘要
Observations of semiconductor multilayers with backscattered electrons in a scanning electron microscope have been used to revisit the concept of resolution of the backscattering imaging mode. With the support of Monte Carlo simulations of beam specimen interaction, it has been possible to achieve the following conclusions. All the backscattered electrons positively contribute to the image formation independently of their trajectories and specimen exit points. The generation volume does not represent in itself a limit to the resolution, which depends only on the beam size and the signal to noise ratio.
引用
收藏
页码:499 / 504
页数:6
相关论文
共 8 条
  • [1] BOSACCHI A, 1994, IN PRESS P S A, V354
  • [2] FRANCHI S, 1990, 12TH P INT C EL MICR, V1, P380
  • [3] RESOLUTION OF SUPERLATTICE STRUCTURES WITH BACKSCATTERED ELECTRONS IN A SCANNING ELECTRON-MICROSCOPE
    MERLI, PG
    NACUCCHI, M
    [J]. ULTRAMICROSCOPY, 1993, 50 (01) : 83 - 93
  • [4] OGURA K, 1991, JEOL NEWS E, V29, P6
  • [5] OGURA K, 1988, 4TH P AS PAC EM C, P165
  • [6] OGURA K, 1988, 46TH P ANN M EL MICR, P204
  • [7] OGURA K, 1990, 12TH P INT C EL MICR, V1, P404
  • [8] WEILS OC, 1992, P INT SCH ELECTRON M, P479