CALIBRATION OF SCINTILLATION DETECTORS FOR MEV CHARGED FUSION PRODUCTS

被引:9
作者
TUSZEWSKI, M [1 ]
ZWEBEN, SJ [1 ]
机构
[1] PRINCETON PLASMA PHYS LAB,PRINCETON,NJ 08544
关键词
D O I
10.1063/1.1143714
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The light output of ZnS scintillators used to detect escaping fusion products in the TFTR Tokamak is studied with 3.5-MeV alpha and 3-MeV proton beams. The emitted light first increases linearly with beam current and then saturates. In all cases investigated, the saturations start at a fairly constant absorbed power density of about 1 mW/cm2. The scintillators have adequate time response up to 50-100 kHz.
引用
收藏
页码:4542 / 4544
页数:3
相关论文
共 4 条
[1]  
BOIVIN R, 1991, PPPL2797 PRINC PLASM
[2]  
LEVSHIN VL, 1964, SOVIET RES LUMINESCE, V23
[3]   CONSTRAINTS ON ESCAPING ALPHA-PARTICLE DETECTORS FOR IGNITED TOKAMAKS [J].
ZWEBEN, SJ ;
BOIVIN, R ;
LIEW, SL ;
OWENS, DK ;
STRACHAN, JD ;
ULRICKSON, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (11) :3505-3508
[4]  
ZWEBEN SJ, 1991, NUCL FUSION, V31, P2219, DOI 10.1088/0029-5515/31/12/001