DIAGNOSIS OF ICD LEAD FAILURE USING CONTINUOUS EVENT MARKER RECORDING

被引:7
作者
HORTON, RP
CANBY, RC
ROMAN, CA
JESSEN, ME
SCHULTE, TJ
PAGE, RL
机构
[1] UNIV TEXAS,SW MED CTR,DEPT SURG,DIV CARDIOTHORAC SURG,DALLAS,TX 75235
[2] CARDIAC PACEMAKERS INC,ST PAUL,MN
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1995年 / 18卷 / 06期
关键词
EVENT MARKER; AMBULATORY MONITOR; OVERSENSING; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR;
D O I
10.1111/j.1540-8159.1995.tb06976.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Stored electrograms enhance the ability to evaluate therapy episodes in the third-generation implantable cardioverter defibrillator. These electrograms are recorded from either the shocking or rate sensing leads, but not both. As a result, differentiation of certain types of sensing abnormalities may be difficult prior to surgical exploration. We present a case of rate sensing lead failure due to an insulation break. Several minutes of recording of the event marker in the laboratory failed to document any abnormal sensing; the diagnosis was made by recording the event marker on a 24-hour continuous (Holter) monitor. The Holter monitor/event marker combination was of substantial diagnostic value and allowed for a more focused surgical evaluation and treatment.
引用
收藏
页码:1331 / 1334
页数:4
相关论文
共 6 条
[1]   LONG-TERM COMPLICATIONS OF IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LEAD SYSTEMS [J].
ALMASSI, GH ;
OLINGER, GN ;
WETHERBEE, JN ;
FEHL, G .
ANNALS OF THORACIC SURGERY, 1993, 55 (04) :888-892
[2]   THE USE OF BEEP-O-GRAMS IN THE ASSESSMENT OF AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR SENSING FUNCTION [J].
BALLAS, SL ;
RASHIDI, R ;
MCALISTER, H ;
CORBELLI, R ;
MCCOWAN, R ;
WILKOFF, BL ;
CASTLE, LW ;
MORANT, VA ;
SIMMONS, TW ;
MALONEY, JD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (11) :1737-1745
[3]  
GRIMM W, 1993, PACE, V18, P218
[4]   EVALUATION OF OUTPATIENTS EXPERIENCING IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SHOCKS ASSOCIATED WITH MINIMAL SYMPTOMS [J].
HAMER, ME ;
CLAIR, WK ;
WILKINSON, WE ;
GREENFIELD, RA ;
PRITCHETT, ELC ;
PAGE, RL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (05) :938-943
[5]   IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR THERAPY IN THE ABSENCE OF SIGNIFICANT SYMPTOMS - RHYTHM DIAGNOSIS AND MANAGEMENT AIDED BY STORED ELECTROGRAM ANALYSIS [J].
HOOK, BG ;
CALLANS, DJ ;
KLEIMAN, RB ;
FLORES, BT ;
MARCHLINSKI, FE .
CIRCULATION, 1993, 87 (06) :1897-1906
[6]   LONG-TERM MULTICENTER EXPERIENCE WITH A 2ND-GENERATION IMPLANTABLE PACEMAKER-DEFIBRILLATOR IN PATIENTS WITH MALIGNANT VENTRICULAR TACHYARRHYTHMIAS [J].
SAKSENA, S ;
POCZOBUTTJOHANOS, M ;
CASTLE, LW ;
FOGOROS, RN ;
ALPERT, BL ;
KRON, J ;
PACIFICO, A ;
GRIFFIN, J ;
RUSKIN, JN ;
KEHOE, RF ;
YEE, R ;
DORIAN, P ;
KERR, CR ;
LUCERI, RM ;
POLISENO, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (03) :490-499