NEW DETECTOR CALIBRATION FOR THE WAVELENGTH RANGE 35-400 NM BASED ON AN ELECTRICAL SUBSTITUTION RADIOMETER

被引:16
作者
LAUFRAMBS, A
KROTH, U
RABUS, H
TEGELER, E
ULM, G
机构
[1] Physikalisch-Technische Bundesanstalt, 10587 Berlin
关键词
D O I
10.1063/1.1145677
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In the radiometry laboratory of the Physikalisch-Technische Bundesanstalt at the Berlin electron storage ring BESSY a new instrumentation has been set up for the calibration of transfer detectors in the wavelength range from 35 to 400 nm with an aspired uncertainty of <1%. The spectral responsivity of a detector is determined by comparison with the known responsivity of an electrical substitution radiometer operated at liquid helium temperature as the primary detector standard. A first calibration of silicon photodiodes was performed in the spectral range above 250 nm. The spectral responsivity determined with 3% uncertainty was found to agree with previous calibrations conducted with lasers as sources. © 1995 American Institute of Physics.