STROKE VOLUME MEASUREMENTS BY ELECTRICAL BIOIMPEDANCE AND ECHOCARDIOGRAPHY IN HEALTHY-VOLUNTEERS

被引:19
作者
HUANG, KC
STODDARD, M
TSUEDA, K
HEINE, MF
THOMAS, MH
WHITE, M
WIEMAN, TJ
机构
[1] UNIV LOUISVILLE,SCH MED,DEPT ANESTHESIOL,LOUISVILLE,KY 40292
[2] UNIV LOUISVILLE,SCH MED,DEPT MED CARDIOL,LOUISVILLE,KY 40292
[3] UNIV LOUISVILLE,SCH MED,DEPT SURG,LOUISVILLE,KY 40292
关键词
D O I
10.1097/00003246-199011000-00018
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
To evaluate the validity of three equations for estimation of thoracic electrical field size in a new bioimpedance algorithm, stroke volume (SV) as calculated by these equations was compared with that calculated by Doppler echocardiography in 48 healthy volunteers, both lean and obese. When the volume of electrically participating tissue was estimated from body height (modified Sramek) or body height corrected for body habitus (Sramek-Bernstein), there was considerable variation between bioimpedance and Doppler stroke volumes. When the volume of electrically participating tissue was estimated from the actual measurement of the height of the thorax and the circumference at the base of the thorax, the variation in SV differences decreased substantially (Sramek equation), although still considerable for clinical use, and there was no relationship between SV thus obtained and body habitus. Analysis of calculated stroke indices derived by our Doppler echocardiographic standard, as compared with values in the literature, revealed a systematic underestimation. We conclude that the original Sramek equation systematically underestimates SV by 15% to 20%, and the modified Sramek and Sramek-Bernstein equations systematically overestimate SV in females by about 15%, but provide SV values in males in the predicted range. Further studies on the current assumption that the electrical field size is a truncated cone may improve precision of the bioimpedance method.
引用
收藏
页码:1274 / 1278
页数:5
相关论文
共 22 条
[1]  
ALEXANDER J K, 1963, Mod Concepts Cardiovasc Dis, V32, P799
[2]   A NEW STROKE VOLUME EQUATION FOR THORACIC ELECTRICAL BIOIMPEDANCE - THEORY AND RATIONALE [J].
BERNSTEIN, DP .
CRITICAL CARE MEDICINE, 1986, 14 (10) :904-909
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   CHANGES IN CARDIAC OUTPUT WITH AGE [J].
BRANDFONBRENER, M ;
LANDOWNE, M ;
SHOCK, NW .
CIRCULATION, 1955, 12 (04) :557-566
[5]   DETERMINATION OF STROKE VOLUME AND CARDIAC-OUTPUT DURING EXERCISE - COMPARISON OF TWO-DIMENSIONAL AND DOPPLER ECHOCARDIOGRAPHY, FICK OXIMETRY, AND THERMODILUTION [J].
CHRISTIE, J ;
SHELDAHL, LM ;
TRISTANI, FE ;
SAGAR, KB ;
PTACIN, MJ ;
WANN, S .
CIRCULATION, 1987, 76 (03) :539-547
[6]   OBESITY AND CARDIAC-FUNCTION [J].
DEDIVITIIS, O ;
FAZIO, S ;
PETITTO, M ;
MADDALENA, G ;
CONTALDO, F ;
MANCINI, M .
CIRCULATION, 1981, 64 (03) :477-482
[7]   MEASUREMENT OF CARDIAC-OUTPUT BY ELECTRICAL IMPEDANCE AT REST AND DURING EXERCISE [J].
DENNISTON, JC ;
MAHER, JT ;
REEVES, JT ;
CRUZ, JC ;
CYMERMAN, A ;
GROVER, RF .
JOURNAL OF APPLIED PHYSIOLOGY, 1976, 40 (01) :91-95
[8]   INFLUENCE OF SAMPLING SITE AND FLOW AREA ON CARDIAC-OUTPUT MEASUREMENTS BY DOPPLER ECHOCARDIOGRAPHY [J].
DITTMANN, H ;
VOELKER, W ;
KARSCH, KR ;
SEIPEL, L .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (04) :818-823
[9]   DETERMINATION OF CARDIAC-OUTPUT BY SINGLE GATED PULSED DOPPLER ECHOCARDIOGRAPHY [J].
FAST, JH ;
VANDENMERKHOF, L ;
BLANS, W ;
VANLEEUWEN, K ;
UIJEN, G .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1988, 21 (01) :33-42
[10]   PREDICTION OF NORMAL BLOOD-VOLUME - RELATION OF BLOOD-VOLUME TO BODY HABITUS [J].
FELDSCHUH, J ;
ENSON, Y .
CIRCULATION, 1977, 56 (04) :605-612