NEUTRON MEASUREMENTS AROUND MEDICAL ELECTRON-ACCELERATORS BY ACTIVE AND PASSIVE DETECTION TECHNIQUES

被引:66
作者
TOSI, G
TORRESIN, A
AGOSTEO, S
PARA, AF
SANGIUST, V
ZENI, L
SILARI, M
机构
[1] POLITECN MILAN,CESNEF,INST INGN NUCL,I-20133 MILAN,ITALY
[2] IST TECNOL BIOMED AVANZATE,CONSIGLIO NAZL RIC,I-20131 MILAN,ITALY
关键词
HIGH-ENERGY ELECTRON ACCELERATORS; PHOTONEUTRONS; HE-3 NEUTRON DETECTORS; FAST NEUTRON DOSIMETRY; RADIOTHERAPY;
D O I
10.1118/1.596751
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Passive and active detection techniques have been employed in order to measure neutron fluence rates and corresponding exposure rates around medical electron accelerators operating at energies well above the neutron binding energies of the structural materials. In these conditions from the treatment head, in the direct photon flux and from the shielded region, a fast neutron flux emerges which is partly absorbed and partly scattered by the walls, eventually establishing a nearly uniform thermal and epithermal flux in the room. Both direct and scattered flux contribute to the dose to the patient. A smaller neutron dose rate can also be found outside the treatment room, where the therapy staff works. Passive detectors, of moderation type, have been employed in the treatment room and He-3 active detectors in the external zones. For the treatment room the activation data were compared with results of Monte Carlo simulation of the neutron transport in the room. Technical features of the two measures are briefly presented and results obtained around three different types of accelerators are reported. At the higher beam energies, i.e., 25 MV, a neutron dose of 0.36 Sv was estimated in the treatment field in addition to a therapeutic x-ray dose of 50 Gy. At lower energies or out of the treatment field the neutron dose drops significantly. In the external zones the dose rates everywhere are below the recommended limits and normally very low, the highest values being recorded in positions very close to the access door of the treatment room.
引用
收藏
页码:54 / 60
页数:7
相关论文
共 17 条
[1]  
AGOSTEO S, 1989, 75 C NAZ SOC IT FIS
[2]  
[Anonymous], 1984, 79 NAT COUNC RAD PRO
[3]   A NEW TYPE OF NEUTRON SPECTROMETER [J].
BRAMBLETT, RL ;
EWING, RI ;
BONNER, TW .
NUCLEAR INSTRUMENTS & METHODS, 1960, 9 (01) :1-12
[4]  
EISENHAUER CM, 1981, 4TH S NEUTR DOS COMM
[5]  
ENGLE WW, 1967, US AEC K1693 REP
[6]  
GUDOWSKA I, 1988, RADIAT PROT DOSIM, V23, P345
[7]  
KONDO I, 1974, RCN74105 REACT CENTR
[8]   TRANSPORT OF ACCELERATOR PRODUCED NEUTRONS IN A CONCRETE ROOM [J].
MCCALL, RC ;
JENKINS, TM ;
SHORE, RA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (01) :1593-1602
[9]  
MCCALL RC, 1981, 4TH S NEUTR DOS COMM
[10]  
MCCALL RC, 1988, RAD PROTECTION PRACT