ARTIFICIAL-VENTILATION FOR CARDIOPULMONARY RESUSCITATION

被引:9
作者
SAINSBURY, DA [1 ]
DAVIS, R [1 ]
WALKER, MC [1 ]
机构
[1] REPATRIAT GEN HOSP, INTENS CARE UNIT, DAW PK, SA 5041, AUSTRALIA
关键词
D O I
10.5694/j.1326-5377.1984.tb132905.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The ability of 100 members of the nursing staff to administer artificial ventilation was tested on a manikin. Six techniques of artificial ventilation were used. A minimum minute volume of 6 l was chosen as the criterion for adequate ventilation. Of the nurses 88% passed using mouth-to-mouth ventilation, 66% passed using mouth-to-face-mask, 55% passed using the RM-1 injector system, 52% passed using the Mistviva injector system, while very low pass rates of 25% and 15% were obtained with the Air-Viva and Mapleson B circuits, respectively. The mouth-to-mouth technique was the best form of ventilatory support provided by the nursing staff tested in this trial. In general, the use of mask systems should be restricted to highly trained staff members who have demonstrated continuing proficiency.