ADEQUACY OF IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR LEAD PLACEMENT FOR TACHYARRHYTHMIA DETECTION BY SINUS RHYTHM ELECTROGRAM AMPLITUDE

被引:8
作者
MICHELSON, BI [1 ]
IGEL, DA [1 ]
WILKOFF, BL [1 ]
机构
[1] CLEVELAND CLIN FDN, CLEVELAND, OH 44195 USA
关键词
D O I
10.1016/S0002-9149(99)80328-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study examines whether the current clinical practice of using a 5 mV minimum amplitude during normal sinus rhythm (NSR) ensures adequate detection during subsequent episodes of ventricular fibrillation (VF) at the time of the implantable cardioverter-defibrillator (ICD) threshold testing. Risk of nondetection occurs with ICDs when a substantial portion of the individual cardiac events on an electrogram goes undetected, Detection risk was assessed by 2 methods: percentage of missed cardiac events (incidence of signal dropout), and mean electrogram amplitude During ICD implantation and testing in 27 patients utilizing 41 lead positions, 135 episodes of Vf were induced and analyzed, On 64 occasions, the countershock was not successful in achieving cardioversion, and the continuing electrical activity was analyzed as a separate group of postshock waveforms, Thresholds of 1 and 2 mV were applied to each individual cardiac depolarization in a VF episode, Significant risk of nondetection was assumed when greater than or equal to 10% of individual events displayed dropout, Underdetection by signal dropout occurred in 11 of 135 preshock arrhythmia signals (8.1%) from 3 patients at a 2 mV threshold, and in 6 of 135 signals (4.4%) at a 1 mV threshold, A mean NSR amplitude greater than or equal to 5 mV was associated with significantly lower risk of non-detection during subsequent VF episodes at both 1 and 2 mV thresholds (largest p<0.001). Similar results were observed in analysis of postshock arrhythmia signals, Further examination of signal dropout and linear regression criteria suggest that in order to eliminate the possibility of nondetection at a 1 mV threshold, minimum NSR amplitudes of 8.5 and 10.0 mV, respectively are required.
引用
收藏
页码:1162 / 1166
页数:5
相关论文
共 14 条
[1]  
AKHTAR M, 1988, CARDIOLOGY, P1
[2]  
BARDY GH, 1988, PACE, V11, P485
[3]  
BEYLER M, 1983, PACE, V6, P314
[4]   CORONARY REVASCULARIZATION AND IMPLANTATION OF THE AUTOMATIC CARDIOVERTER DEFIBRILLATOR - RELIABILITY OF IMMEDIATE INTRAOPERATIVE TESTING [J].
BLAKEMAN, BM ;
PIFARRE, R ;
SCANLON, PJ ;
WILBER, DJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :86-91
[5]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - CLINICAL-EXPERIENCE, COMPLICATIONS, AND FOLLOW-UP IN 25 PATIENTS [J].
BORBOLA, J ;
DENES, P ;
EZRI, MD ;
HAUSER, RG ;
SERRY, C ;
GOLDIN, MD .
ARCHIVES OF INTERNAL MEDICINE, 1988, 148 (01) :70-76
[6]   IMPLANTATION OF THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (AICD) - PRACTICAL ASPECTS [J].
CANNOM, DS ;
WINKLE, RA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06) :793-809
[7]   MEASUREMENT OF VENTRICULAR ELECTROGRAM AMPLITUDE DURING INTRAOPERATIVE INDUCTION OF VENTRICULAR TACHYARRHYTHMIAS [J].
ELLENBOGEN, KA ;
WOOD, MA ;
STAMBLER, BS ;
WELCH, WJ ;
DAMIANO, RJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (11) :1017-1022
[8]   CORRELATION BETWEEN THE VENTRICULAR ELECTROGRAM AMPLITUDE IN SINUS RHYTHM AND IN VENTRICULAR-FIBRILLATION [J].
LEITCH, JW ;
YEE, R ;
KLEIN, GJ ;
JONES, DL ;
MURDOCK, CJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (09) :1105-1109
[9]  
PANNIZZO F, 1986, PACE, V9, P304
[10]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - T-WAVE SENSING IN THE NEWEST GENERATION [J].
SINGER, I ;
DEBORDE, R ;
VELTRI, EP ;
SIDDOWAY, LA ;
LEVINE, JH ;
GRIFFITH, LSC ;
GUARNIERI, T .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1584-1591