TRANSESOPHAGEAL ECHOCARDIOGRAPHY DURING RADIOFREQUENCY CATHETER ABLATION OF VENTRICULAR-TACHYCARDIA

被引:30
作者
SAXON, LA
STEVENSON, WG
FONAROW, GC
MIDDLEKAUFF, HR
YEATMAN, LA
SHERMAN, CT
CHILD, JS
机构
[1] Division of Cardiology, Department of Medicine, UCLA Medical Center, Los Angeles, CA
关键词
D O I
10.1016/0002-9149(93)90880-L
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Radiofrequency lesion formation requires stable catheter tip/endocardial contact. Energy delivery is limited when temperatures are > 100-degrees-C due to coagulum formation at the catheter tip. Transesophageal echocardiographic imaging may be useful for monitoring catheter position and detecting boiling. Transesophageal echocardiographic images were recorded during production of 22 radiofrequency lesions in bovine myocardium in a saline bath. Lesion size, tissue temperature and appearance of echo contrast (bubbles) were assessed. In 11 patients, transesophageal echocardiography was used to guide catheter movement and detect boiling during radiofrequency ablation for ventricular tachycardia. In the tissue bath, the appearance of echo bubbles was associated with visual bubbling at the catheter tip, tissue temperatures > 60-degrees-C and larger lesions (284 +/- 165 vs 30 +/- 54 mm3; p < 0.001). In humans, transesophageal images easily identified the catheter tip in either ventricle and enabled continuous observation of electrode-tissue contact during radiofrequency application. Transesophageal echocardiographic bubbles appeared in 59 of 217 radiofrequency applications (27%). Continued radiofrequency application after appearance of bubbles was followed by an increase in impedance. Prolonged placement of the probe in heavily sedated patients resulted in a mild sore throat, but no other complications. Transesophageal echocardiographic imaging enables continuous monitoring of catheter position during radiofrequency energy application. The abrupt appearance of echo bubbles indicates boiling and impending coagulum formation at the catheter tip.
引用
收藏
页码:658 / 661
页数:4
相关论文
共 14 条
[1]  
BOMMER WJ, 1984, J AM COLL CARDIOL, V3, P6, DOI 10.1016/S0735-1097(84)80423-4
[2]   DIAGNOSIS AND CURE OF THE WOLFF-PARKINSON-WHITE SYNDROME OR PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIAS DURING A SINGLE ELECTROPHYSIOLOGIC TEST [J].
CALKINS, H ;
SOUSA, J ;
ELATASSI, R ;
ROSENHECK, S ;
DEBUITLEIR, M ;
KOU, WH ;
KADISH, AH ;
LANGBERG, JJ ;
MORADY, F .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (23) :1612-1618
[3]  
FEIGENBAUM H, 1986, ECHOCARDIOGR-J CARD, P150
[4]   CATHETER ABLATION OF CANINE MYOCARDIUM WITH RADIOFREQUENCY ENERGY [J].
FRANKLIN, JO ;
LANGBERG, JJ ;
OEFF, M ;
FINKBEINER, WE ;
HERRE, JM ;
GRIFFIN, JC ;
SCHEINMAN, MM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :170-176
[5]   TRANSESOPHAGEAL ECHOCARDIOGRAPHY TO IMPROVE POSITIONING OF RADIOFREQUENCY ABLATION CATHETERS IN LEFT-SIDED WOLFF-PARKINSON-WHITE SYNDROME [J].
GOLDMAN, AP ;
IRWIN, JM ;
GLOVER, MU ;
MICK, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (08) :1245-1250
[6]  
HAINES D E, 1991, Journal of Cardiovascular Electrophysiology, V2, P509, DOI 10.1111/j.1540-8167.1991.tb01353.x
[7]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[8]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[9]   CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR PATHWAYS (WOLFF-PARKINSON-WHITE SYNDROME) BY RADIOFREQUENCY CURRENT [J].
JACKMAN, WM ;
WANG, XZ ;
FRIDAY, KJ ;
ROMAN, CA ;
MOULTON, KP ;
BECKMAN, KJ ;
MCCLELLAND, JH ;
TWIDALE, N ;
HAZLITT, HA ;
PRIOR, MI ;
MARGOLIS, PD ;
CALAME, JD ;
OVERHOLT, ED ;
LAZZARA, R .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (23) :1605-1611
[10]   CATHETER ABLATION OF ATRIOVENTRICULAR JUNCTION USING RADIOFREQUENCY CURRENT IN 17 PATIENTS - COMPARISON OF STANDARD AND LARGE-TIP CATHETER ELECTRODES [J].
JACKMAN, WM ;
WANG, XZ ;
FRIDAY, KJ ;
FITZGERALD, DM ;
ROMAN, C ;
MOULTON, K ;
MARGOLIS, PD ;
BOWMAN, AJ ;
KUCK, KH ;
NACCARELLI, GV ;
PITHA, JV ;
DYER, J ;
LAZZARA, R .
CIRCULATION, 1991, 83 (05) :1562-1576