DETERMINATION OF CLOCAPRAMINE AND ITS METABOLITES IN PLASMA BY AUTOMATED COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY

被引:4
作者
HIKIDA, K [1 ]
INOUE, Y [1 ]
KOJIMA, N [1 ]
OHKURA, Y [1 ]
机构
[1] KYUSHU UNIV 62,FAC PHARMACEUT,FUKUOKA 812,JAPAN
关键词
bovine serum albumin-coated cyanopropyl silica gel column; clocapramine; clospipramine; Column-switching high performance liquid chromatography; dehydroclospipramine; metabolite; plasma;
D O I
10.2116/analsci.6.367
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated column-switching high performance liquid chromatographic method with ultraviolet detection is described for rapid and sensitive determination of clocapramine, a psychotropic drug, and its two metabolites, clospipramine and dehydroclospipramine, in dog and human plasmas. The drug and the metabolites were extracted together with trifluperidol as an internal standard with a heptane-chloroform mixture from plasma. The extract was subjected to column-switching high performance liquid chromatography using a bovine serum albumin-coated cyanopropyl silica gel column, a newly developed sample clean-up column, and a reversed phase column (TSK gel ODS-80TM) for analytical separation. The detection limits (S/N=3) of clocapramine, clospipramine and dehydroclospipramine were 0.5 - 1.0 ng ml-1. © 1990, The Japan Society for Analytical Chemistry. All rights reserved.
引用
收藏
页码:367 / 370
页数:4
相关论文
共 9 条
[1]  
HIKIDA K, 1989, BUNSEKI KAGAKU, V38, P135
[2]  
HIKIDA K, 1988, BUNSEKI KAGAKU, V37, P566
[3]   DETERMINATION OF BROMPERIDOL IN SERUM BY AUTOMATED COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
HIKIDA, K ;
INOUE, Y ;
MIYAZAKI, T ;
KOJIMA, N ;
OHKURA, Y .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1989, 495 :227-234
[4]  
HIKIDA K, 1989, JPN J CLIN CHEM, V18, P180
[5]  
IMAI H, 1989, ANAL SCI, V389, P5
[6]   PHARMACOKINETIC STUDY OF IMINODIBENZYL ANTIPSYCHOTIC-DRUGS, CLOCAPRAMINE AND Y-516 IN DOG AND MAN [J].
ISHIGOOKA, J ;
MURASAKI, M ;
WAKATABE, H ;
MIURA, S ;
HIKIDA, K ;
SHIBATA, M ;
NOBUNAGA, H .
PSYCHOPHARMACOLOGY, 1989, 97 (03) :303-308
[7]   SYSTEMATIC-APPROACH TO COLUMN SWITCHING [J].
RAMSTEINER, KA .
JOURNAL OF CHROMATOGRAPHY, 1988, 456 (01) :3-20
[8]  
SADANAGA T, 1980, JPN J CLIN CHEM, V9, P351
[9]  
YOSHIDA H, 1982, CHEM PHARM BULL, V30, P2287