COMPARISON OF INTRABRACHIAL AND FINGER BLOOD-PRESSURE IN HEALTHY ELDERLY VOLUNTEERS

被引:75
作者
RONGEN, GA
BOS, WJW
LENDERS, JWM
VANMONTFRANS, GA
VANLIER, HJJ
VANGOUDOEVER, J
WESSELING, KH
THIEN, T
机构
[1] Departments of Medicine, Division of General Internal Medicine, University Hospital, Nijmegen
[2] Department of Medicine, University Hospital, Institute of Biomedical Research and Academic Medical Centre, Amsterdam
关键词
FINAPRES; INTRAARTERIAL PRESSURE; HEALTHY ELDERLY; HEAD-UP TILT; STANDING; VALSALVA; MENTAL ARITHMETIC;
D O I
10.1016/0895-7061(94)00000-2
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
This study was performed to compare continuous Finapres (FIN) and intrabrachial (IAP) blood pressure in healthy elderly volunteers. Fifteen elderly subjects (age 71 to 83) without cardiovascular disease and an intraarterial mean (range) systolic and diastolic blood pressure of 162 (122 to 195) and 73 (62 to 88) mm Hg, respectively, participated in the study. A 10-min head-up tilt, 10 min active standing, a 15-sec Valsalva, and a 5-min mental arithmetic were performed in random order. Beat-to-beat values of systolic, diastolic, and mean arterial pressure were analyzed. At rest, FIN underestimated IAP by 16.8 +/- 2.6 (SE), 10.8 +/- 1.5, and 17.5 +/- 1.6 mm Hg for systolic, diastolic, and mean arterial blood pressure, respectively (P < .05). During head-up tilt, FIN overestimated the intraarterial systolic blood pressure response by 7.2 +/- 1.6 (SE) mm Hg (P < .05). Group-averaged changes in diastolic and mean arterial IAP were followed closely by FIN. During standing, Finapres closely followed intraarterial diastolic and mean arterial pressure but the increase in systolic blood pressure was higher at the finger as compared to intrabrachial recordings, resembling the results of head-up tilt. During the Valsalva maneuver, maximal responses in systolic, diastolic, and mean arterial pressure were underestimated by FIN by 12.1 +/- 3.3 (SE), 6.8 +/- 2.7, and 7.1 +/- 1.7 mm Hg, respectively (P < .05 for all parameters). During mental arithmetic, FIN underestimated the intraarterial systolic blood pressure response by 6.1 +/- 2.7 (SE) mm Hg (P < .05), while diastolic and mean arterial pressure responses were followed correctly by FIN. It is concluded that apart from systolic blood pressure, FIN closely follows intraarterial blood pressure responses for the orthostatic maneuvers and mental arithmetic. During Valsalva, the rapid changes in blood pressure were followed in direction but not in magnitude.
引用
收藏
页码:237 / 248
页数:12
相关论文
共 29 条
[1]  
Wieling W., Non-invasive continuous recording of heart rate and blood pressure in the evaluation of neu-rocardiovascular control, A Textbook of Clinical Disorders of the Autonomic Nervous System, pp. 291-311, (1992)
[2]  
Robbe H.W., Mulder L.J., Ruddel H., Et al., Assessment of baroreceptor reflex sensitivity by means of spectral analysis, Hypertension, 10, pp. 538-543, (1987)
[3]  
Parati G., Di-Rienzo M., Bertinieri G., Et al., Evaluation of the baroreceptor-heart rate reflex by 24-hour intra-arterial blood pressure monitoring in humans, Hypertension, 12, pp. 214-222, (1988)
[4]  
Penaz J., Photoelectric measurement of blood pressure, volume and flow in the finger, in Digest 10 in the Conf Med Biol Engng, Dresden, (1973)
[5]  
Wesseling K.H., De Wit B., Settels J.J., Et al., On the indirect registration of finger blood pressure after Penaz, Funkt Biol Med, 1, pp. 245-250, (1982)
[6]  
Wesseling K.H., Settels J.J., De Wit B., The measurement of continuous finger arterial pressure noninvasively in stationary subjects, Biological and Psychological Factors in Cardiovascular Disease, pp. 355-375, (1986)
[7]  
Imholz B.P., Van-Montfrans G.A., Settels J.J., Et al., Continuous non-invasive blood pressure monitoring: Reliability of Finapres device during the Valsalva manoeuvre, Cardiovasc Res, 22, pp. 390-397, (1988)
[8]  
Parati G., Casadei R., Groppelli A., Et al., Comparison of finger and intra-arterial blood pressure monitoring at rest and during laboratory testing, Hypertension, 13, pp. 647-655, (1989)
[9]  
Imholz B.P., Settels J.J., Van-Der-Meiracker A.H., Et al., Non-invasive continuous finger blood pressure measurement during orthostatic stress compared to intra-arterial pressure, Cardiovasc Res, 24, pp. 214-221, (1990)
[10]  
Gorback M.S., Quill T.J., Lavine M.L., The relative accuracies of two automated noninvasive arterial pressure measurement devices, J Clin Monit, 7, pp. 13-22, (1991)