PREPARATION AND STABILITY-TESTS FOR POLYACRYLAMIDE-COATED CAPILLARIES FOR CAPILLARY ELECTROPHORESIS

被引:46
作者
ENGELHARDT, H
CUNATWALTER, MA
机构
[1] Instrumentelle Analytik, Umweltanalytik, Universität des Saarlandes
关键词
D O I
10.1016/0021-9673(95)00480-B
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The stability of polyacrylamide coatings at high pH values has been tested. It could be shown that overall stability is not limited by the coating of the fused-silica surface with organo silanes or Grignard reagents. The instability of polyacrylamide capillaries could be traced back to the hydrolysis of the amide bond at pH 10. The Si-O-Si surface bonds are not affected at this pH by hydrolysis. The application of the sol-gel process to the preparation of stable homogeneous sublayers and a high density of olefinic groups within the capillary is described, The advantage of this method is the simple coating reaction and its independence of fused-silica surface properties.
引用
收藏
页码:27 / 33
页数:7
相关论文
共 20 条