OPTIMIZATION OF NMR RECEIVER BANDWIDTH BY INDUCTIVE COUPLING

被引:40
作者
RAAD, A
DARRASSE, L
机构
[1] Université de PARIS-SUD, Institut d'Electronique Fondamentale, CNRS UA 22, 91405 Orsay, bât
关键词
NMR COIL; INDUCTIVE COUPLING; RECEIVER BANDWIDTH; SIGNAL-TO-NOISE RATIO; FAST MRI; LOW-FIELD MRI;
D O I
10.1016/0730-725X(92)90373-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We show, by theory and experiment, that inductive coupling can be used in overcoupled mode to widen the bandwidth of a high-Q NMR coil with only a negligible degradation of signal-to-noise ratio over the bandwidth of interest. The receiver bandwidth depends on the coupling coefficient between the NMR coil and the coupling coil rather than on the quality factor of the NMR coil alone. The overall bandwidth can be optimized by a judicious choice of the coupling coefficient. Moreover, this method permits wireless reception without the need for returning and rematching despite changes of NMR coils or samples. This technique has been incorporated in a 0.1-T imager for clinical routine. It achieves a typical bandwidth five times greater than that using a classic 50-OMEGA matching method.
引用
收藏
页码:55 / 65
页数:11
相关论文
共 17 条
  • [1] Decorps, Blondet, Reutenauer, Albrand, Remy, An inductively coupled, series-tuned NMR probe, J. Magn. Reson., 65, pp. 100-109, (1985)
  • [2] Froncisz, Jesmanowicz, Hyde, Inductive (flux linkage) coupling to local coils in magnetic resonance imaging and spectroscopy, J. Magn. Reson., 66, pp. 135-143, (1986)
  • [3] Schnall, Barlow, Subramanian, Leigh, Wireless implanted magnetic resonance probes for in vivo NMR, J. Magn. Reson., 68, pp. 161-167, (1986)
  • [4] Smith, Pye, Ridjway, Kean, Douglas, Rowan, Best, Surface coils for vertical field MRI system, Montreux: Second congress of the ESMRMB proceedings, (1985)
  • [5] Darrasse, Raad, Boudou, Guillemot, Varga, Optimization of receiver coil bandwidth by inductive coupling, New York: Proceedings of the 9th Annual Meeting of the Society of Magnetic Resonance in Medicine, (1990)
  • [6] Rzedzian, Pykett, Instant images of the human heart using a new whole-body MR imaging system, AJR, 149, pp. 245-250, (1987)
  • [7] Hoult, Lauterbur, The sensitivity of the zeugmatographic experiment involving human samples, J. Magn. Reson., 34, pp. 425-433, (1979)
  • [8] Ott, Noise Reduction Techniques in Electronic Systems, (1976)
  • [9] Hoult, Richards, The signal-to-noise ratio of the nuclear magnetic resonance experiment, J. Magn. Reson., 24, pp. 71-85, (1976)
  • [10] Terman, Radio Engineer's Handbook, pp. 150-155, (1943)