LONG-TERM EEG-VIDEO-AUDIO MONITORING - COMPUTER-DETECTION OF FOCAL EEG SEIZURE PATTERNS

被引:61
作者
PAURI, F
PIERELLI, F
CHATRIAN, GE
ERDLY, WW
机构
[1] UNIV WASHINGTON,MED CTR,SCH MED,DEPT MED,DIV NEUROL,EEG & CLIN NEUROPHYSIOL SECT,RC-97,SEATTLE,WA 98195
[2] UNIV WASHINGTON,SCH MED,DEPT NEUROL SURG,SEATTLE,WA 98195
[3] UNIV WASHINGTON,SCH ARTS & SCI,DEPT PSYCHOL,SEATTLE,WA 98195
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1992年 / 82卷 / 01期
关键词
EEG-VIDEO-AUDIO MONITORING; FOCAL EEG SEIZURES; PARTIAL SEIZURES; COMPUTER SEIZURE DETECTION; VISUAL SEIZURE DETECTION; SEIZURE SIGNALING;
D O I
10.1016/0013-4694(92)90175-H
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Twelve individuals with medically refractory partial seizures had undergone EEG-video-audio (EVA) monitoring over 1-15 (mean 10.5) days. We selectively reexamined available 15-channel EEGs (video-cassettes) totaling 461 h and containing 253 EEG focal seizures. Computer analysis (CA) of these bipolar records was performed using a mimetic method of seizure detection at 6 successive computer settings. We determined the computer parameters at which this method correctly detected a reasonably large percentage of seizures (81.42%) while generating an acceptable rate of false positive results (5.38/h). These parameters were adopted as the default setting for identifying focal EEG seizure patterns in all subsequent long-term bipolar scalp and sphenoidal recordings. Factors hindering or facilitating automatic seizure identification are discussed. It is concluded that on-line computer detection of focal EEG seizure patterns by this method offers a satisfactory alternative to and represents a distinct improvement over the extremely time consuming and fatiguing off-line fast visual review (FVR). Combining CA with seizure signaling (SS) by the patients and other observers increased the correct detections to 85.38%. CA is best used in conjunction with SS.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 10 条
[1]   AUTOMATIC RECOGNITION AND QUANTIFICATION OF INTERICTAL EPILEPTIC ACTIVITY IN HUMAN SCALP EEG [J].
GOTMAN, J ;
GLOOR, P .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1976, 41 (05) :513-529
[2]   AUTOMATIC RECOGNITION OF EPILEPTIC SEIZURES IN THE EEG [J].
GOTMAN, J .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1982, 54 (05) :530-540
[3]   AUTOMATIC SEIZURE DETECTION - IMPROVEMENTS AND EVALUATION [J].
GOTMAN, J .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1990, 76 (04) :317-324
[4]  
GOTMAN J, 1986, ADV NEUROL, V46, P249
[5]  
GOTMAN J, 1985, ELECTROENCEPH CLIN S, V37
[6]   2-PULSE MEASURES OF TEMPORAL INTEGRATION IN THE FOVEA AND PERIPHERAL RETINA [J].
GOTTLIEB, MD ;
KIETZMAN, ML ;
BERENHAUS, IJ .
PERCEPTION & PSYCHOPHYSICS, 1985, 37 (02) :135-138
[7]  
GUMNIT RJ, 1987, ADV NEUROLOGY, V46
[8]  
PAURI F, 1990, Neurophysiologie Clinique, V20, p95S
[9]   LONG-TERM EEG-VIDEO-AUDIO MONITORING - DETECTION OF PARTIAL EPILEPTIC SEIZURES AND PSYCHOGENIC EPISODES BY 24-HOUR EEG RECORD REVIEW [J].
PIERELLI, F ;
CHATRIAN, GE ;
ERDLY, WW ;
SWANSON, PD .
EPILEPSIA, 1989, 30 (05) :513-523
[10]  
1985, EPILEPSIA, V26, P268