A CLINICAL COMPARISON OF THE TENDENCY TO CAPSULAR CONTRACTURE BETWEEN SMOOTH AND TEXTURED GEL-FILLED SILICONE MAMMARY IMPLANTS

被引:134
作者
HAKELIUS, L
OHLSEN, L
机构
[1] Department of Plastic and Hand Surgery University Hospital, Uppsala
关键词
D O I
10.1097/00006534-199290020-00012
中图分类号
R61 [外科手术学];
学科分类号
摘要
The aim of this prospective, controlled clinical investigation was to find out if there is a difference in the capsular contracture rate between silicone implants with a smooth or textured surface as the only difference. Twenty-five women with bilateral mammary hypoplasia underwent mammary augmentation. All got a textured implant on one side and a smooth implant on the other. The implants were placed subglandularly. Follow-up examinations were done on six occasions. Three parameters were used for estimation of the tendency to capsular contracture: (1) the patient's opinion on differences in hardness of the breasts, (2) the investigator's classification of capsular contracture, and (3) applanation tonometry. At the end of the follow-up period, after 1 year, all parameters showed with no doubt that the breasts augmented with textured implants had a lower tendency to develop contracting capsules than the breasts augmented with smooth implants.
引用
收藏
页码:247 / 254
页数:8
相关论文
共 29 条
[1]   FURTHER STUDIES ON NATURAL-Y BREAST PROSTHESIS [J].
ASHLEY, FL .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1972, 49 (04) :414-&
[2]   A NEW TYPE OF BREAST PROSTHESIS - PRELIMINARY REPORT [J].
ASHLEY, FL .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1970, 45 (05) :421-&
[3]  
ASPLUND O, 1984, PLAST RECONSTR SURG, V73, P270, DOI 10.1097/00006534-198402000-00022
[4]   OCCURRENCE AND ACTIVITY OF MYOFIBROBLASTS IN HUMAN CAPSULAR TISSUE SURROUNDING MAMMARY IMPLANTS [J].
BAKER, JL ;
CHANDLER, ML ;
LEVIER, RR .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1981, 68 (06) :905-912
[5]   POSITIVE IDENTIFICATION OF SILICONE IN HUMAN MAMMARY CAPSULAR TISSUE [J].
BAKER, JL ;
LEVIER, RR ;
SPIELVOGEL, DE .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1982, 69 (01) :56-60
[6]  
BAKER JL, 1978, S AESTHETIC SURGERY
[7]   BLEEDING OF SILICONE FROM BAG-GEL BREAST IMPLANTS, AND ITS CLINICAL RELATION TO FIBROUS CAPSULE REACTION [J].
BARKER, DE ;
RETSKY, MI ;
SCHULTZ, S .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1978, 61 (06) :836-841
[8]   SURGICAL-CORRECTION OF BREAST DEFORMITIES FOLLOWING LONG-LASTING COMPLICATIONS OF POLYURETHANE-COVERED IMPLANTS [J].
BERRINO, P ;
FRANCHELLI, S ;
SANTI, P .
ANNALS OF PLASTIC SURGERY, 1990, 24 (06) :481-488
[9]   LONG-LASTING COMPLICATIONS WITH THE USE OF POLYURETHANE-COVERED BREAST IMPLANTS [J].
BERRINO, P ;
GALLI, A ;
RAINERO, ML ;
SANTI, PL .
BRITISH JOURNAL OF PLASTIC SURGERY, 1986, 39 (04) :549-553