MAGNETIC-RESONANCE-IMAGING - SPINAL-CORD IMAGING WITH THE TURBO-FLUID ATTENUATED INVERSION-RECOVERY (FLAIR) PULSE SEQUENCE

被引:18
作者
HAJNAL, JV
KASUBOSKI, L
DESOUZA, NM
BYDDER, GM
机构
[1] HAMMERSMITH HOSP,ROYAL POSTGRAD MED SCH,ROBERT STEINER MAGNET RESONANCE UNIT,LONDON W12 0HS,ENGLAND
[2] PICKER INT,BOREHAMWOOD,HERTS,ENGLAND
[3] PICKER INT,HIGHLAND HTS,OH
基金
英国医学研究理事会;
关键词
D O I
10.1016/S0009-9260(05)82956-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Twelve patients with suspected or proven spinal cord disease were studied with a fluid attenuated repeated echo sequence employing a fast spin-echo data acquisition (Turbo-FLAIR) and the results were compared with those obtained with conventional T-1- and T-2- weighted spin-echo sequences. The Turbo-FLAIR sequence utilized an unselected 180 degrees pulse to reduce the signal from CSF followed by a spin-echo sequence based on the Rapid Acquisition with Relaxation Enhancement (RARE) pulse sequence for data acquisition. Three and four echo versions were used. Four low signal intensity lesions (cysts or syrinxes) were seen of which three were better seen with the Turbo-FLAIR sequence than with the other sequences, The remaining lesion was equally web seen. Of the seven high signal lesions found, six were better seen with the Turbo-FLAIR sequence and one was equally well seen. The Turbo-FLAIR sequence is more rapid than the T-2-weighted spin-echo sequence and compares favourably with the combined results of both the T-1- and T-2-weighted spin-echo sequences for lesion conspicuity.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 5 条
[1]   FACTORS INFLUENCING CONTRAST IN FAST SPIN-ECHO MR IMAGING [J].
CONSTABLE, RT ;
ANDERSON, AW ;
ZHONG, J ;
GORE, JC .
MAGNETIC RESONANCE IMAGING, 1992, 10 (04) :497-511
[2]  
DECOENE B, 1992, AM J NEURORADIOL, V13, P1555
[3]   RARE IMAGING - A FAST IMAGING METHOD FOR CLINICAL MR [J].
HENNIG, J ;
NAUERTH, A ;
FRIEDBURG, H .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) :823-833
[4]   FAST SPIN-ECHO MR IMAGING OF THE BRAIN AND SPINE - CURRENT CONCEPTS [J].
JONES, KM ;
MULKERN, RV ;
SCHWARTZ, RB ;
OSHIO, K ;
BARNES, PD ;
JOLESZ, FA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 158 (06) :1313-1320
[5]   USE OF FLUID-ATTENUATED INVERSION-RECOVERY PULSE SEQUENCES FOR IMAGING THE SPINAL-CORD [J].
WHITE, SJ ;
HAJNAL, JV ;
YOUNG, IR ;
BYDDER, GM .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) :153-162