TECHNICAL ASPECTS OF MYOCARDIAL SPECT IMAGING WITH TC-99M SESTAMIBI

被引:170
作者
GARCIA, EV
COOKE, CD
VANTRAIN, KF
FOLKS, R
PEIFER, J
DEPUEY, EG
MADDAHI, J
ALAZRAKI, N
GALT, J
EZQUERRA, N
ZIFFER, J
AREEDA, J
BERMAN, DS
机构
[1] EMORY UNIV, SCH MED, ATLANTA, GA 30322 USA
[2] CEDARS SINAI MED CTR, LOS ANGELES, CA 90048 USA
[3] GEORGIA INST TECHNOL, ATLANTA, GA 30332 USA
[4] ST LUKES HOSP, NEW YORK, NY USA
关键词
D O I
10.1016/0002-9149(90)90608-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Most reports to date using single photon emission computed tomography (SPECT) with technetium-99m (Tc-99m) sestamibi have used acquisition parameters that were optimized for thallium-201. To fully utilize the superior imaging characteristics of Tc-99m sestamibi, there is a need to optimize the technical aspects of SPECT imaging for this agent. Performance can be enhanced through the careful selection of optimal radiopharmaceutical doses, imaging sequences, acquisition parameters, reconstruction filters, perfusion quantification methods and multidimensional methods for visualizing perfusion distribution. The current report describes theoretical considerations, phantom studies and preliminary patient results that have led to optimized protocols, developed at Emory University and Cedars-Sinai Medical Center, for same-day reststress studies, given existing instrumentation and recommended dose limits. The optimizations were designed to fit a low-dose-high-dose rest-stress same-day imaging protocol. A principal change in the acquisition parameters compared with previous Tc-99m sestamibi protocols is the use of a highresolution collimator. The approach is being developed in both prone and supine positions. A new method for extracting a 3-dimensional myocardial count distribution has been developed that uses spherical coordinates to sample the apical region and cylindrical coordinates to sample the rest of the myocardium. New methods for visualizing the myocardial distribution in multiple dimensions are also described, with improved 2-dimensional, as well as 3- and 4-dimensional (3 dimensions plus time) displays. In the improved 2-dimensional display, distance-weighted and volume-weighted polar maps are used that appear to significantly improve the representation of defect location and defect extent, respectively. The protocols established and the methods developed for acquisition, quantification and display of Tc-99m sestamibi SPECT should provide improved image quality and accuracy, compared to current thallium-201 and Tc-99m sestamibi studies reported to date. © 1990.
引用
收藏
页码:E23 / E31
页数:9
相关论文
共 13 条
  • [1] COOKE CD, 1990, 1ST C VIS BIOM COMP, P185
  • [2] QUANTITATIVE ROTATIONAL TL-201 TOMOGRAPHY FOR IDENTIFYING AND LOCALIZING CORONARY-ARTERY DISEASE
    DEPASQUALE, EE
    NODY, AC
    DEPUEY, EG
    GARCIA, EV
    PILCHER, G
    BREDLAU, C
    ROUBIN, G
    GOBER, A
    GRUENTZIG, A
    DAMATO, P
    BERGER, HJ
    [J]. CIRCULATION, 1988, 77 (02) : 316 - 327
  • [3] ESQUERRE JP, 1989, J NUCL MED, V30, P398
  • [4] FOLKS R, 1989, Journal of Nuclear Medicine, V30, P795
  • [5] GALT J R, 1988, Journal of Nuclear Medicine, V29, P797
  • [6] EFFECTS OF MYOCARDIAL WALL THICKNESS ON SPECT QUANTIFICATION
    GALT, JR
    GARCIA, EV
    ROBBINS, WL
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (02) : 144 - 150
  • [7] GARCIA EV, 1985, J NUCL MED, V26, P17
  • [8] KHAN JK, 1988, J AM COLL CARDIOL, V11, pA31
  • [9] KING MA, 1988, J NUCL MED, V29, P1980
  • [10] MARCASSA C, 1990, J NUCL MED, V31, P173