DETERMINATION OF A CEPHALOSPORIN ANTIBIOTIC, CEFTIBUTEN, IN HUMAN PLASMA WITH COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ULTRAVIOLET DETECTION

被引:12
作者
PAN, HT [1 ]
KUMARI, P [1 ]
LIM, J [1 ]
LIN, CC [1 ]
机构
[1] SCHERING PLOUGH CORP,DEPT DRUG METAB,BLOOMFIELD,NJ 07003
关键词
D O I
10.1002/jps.2600810715
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
A column-switching high-performance liquid chromatographic assay combined with a heart-cutting technique and UV detection (LC/LC/UV) was developed to determine ceftibuten, a new oral cephalosporin, in human plasma. Plasma samples were directly injected into the first chromatographic column for samples cleanup and extraction. Thereafter, a switching valve, located at the junction of the first cleanup column and the second analytical column, opened during a 4-6-min interval after injection to transfer the segment containing ceftibuten into the analytical column for quantitation. The method was used routinely in pharmacokinetic studies and has the advantages of simplicity, improved sensitivity, and selectivity. Concentrations of ceftibuten in plasma over a range of 0.1-20-mu-g/mL can be determined with high precision and reproducibility. A single, oral, 200-mg dose of ceftibuten in humans resulted in a maximum concentration of 9.79-mu-g/mL, an area under the plot of concentration of drug in plasma versus time (0-12 h) of 47.77-mu-g.h/mL, and a half-life of 3.2 h. These results document the utility of the LC/LC/UV method in clinical pharmacokinetic studies.
引用
收藏
页码:663 / 666
页数:4
相关论文
共 9 条
[1]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC COLUMN SWITCHING TECHNIQUES FOR RAPID HYDROCARBON GROUP-TYPE SEPARATIONS [J].
ALFREDSON, TV .
JOURNAL OF CHROMATOGRAPHY, 1981, 218 (1-3) :715-728
[3]   COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC (HPLC) DETERMINATION OF HYDROCHLOROTHIAZIDE IN RAT, DOG, AND HUMAN PLASMA [J].
KUO, BS ;
MANDAGERE, A ;
OSBORNE, DR ;
HWANG, KK .
PHARMACEUTICAL RESEARCH, 1990, 7 (12) :1257-1261
[4]   COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR DETERMINATION OF A NEW ANTIVIRAL AGENT, CYCLARADINE, IN SERUM [J].
LIM, JM ;
LIN, CC .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1986, 29 (06) :977-979
[5]   OPTIMIZATION OF MULTIDIMENSIONAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FOR THE DETERMINATION OF DRUGS IN PLASMA BY DIRECT INJECTION, MICELLAR CLEANUP AND PHOTODIODE ARRAY DETECTION [J].
POSLUSZNY, JV ;
WEINBERGER, R ;
WOOLF, E .
JOURNAL OF CHROMATOGRAPHY, 1990, 507 :267-276
[6]   DETERMINATION OF DRUG SUBSTANCES IN BIOLOGICAL-FLUIDS BY DIRECT INJECTION MULTIDIMENSIONAL LIQUID-CHROMATOGRAPHY WITH A MICELLAR CLEANUP AND REVERSED-PHASE CHROMATOGRAPHY [J].
POSLUSZNY, JV ;
WEINBERGER, R .
ANALYTICAL CHEMISTRY, 1988, 60 (18) :1953-1958
[7]  
ROSTON DA, 1988, ANAL CHEM, V60, P950
[8]   ONLINE MULTIDIMENSIONAL LIQUID-CHROMATOGRAPHIC DETERMINATION OF POLYNUCLEAR AROMATIC-HYDROCARBONS IN COMPLEX SAMPLES [J].
SONNEFELD, WJ ;
ZOLLER, WH ;
MAY, WE ;
WISE, SA .
ANALYTICAL CHEMISTRY, 1982, 54 (04) :723-727
[9]  
VEALS JW, 1988, AM LAB, V20, P42