THE AMS 700 INFLATABLE PENILE PROSTHESIS - LONG-TERM EXPERIENCE WITH THE CONTROLLED EXPANSION CYLINDERS

被引:47
作者
QUESADA, ET [1 ]
LIGHT, JK [1 ]
机构
[1] BAYLOR COLL MED,SCOTT DEPT UROL,HOUSTON,TX 77030
关键词
PENIS; PENILE ERECTION; IMPOTENCE; PENILE PROSTHESIS;
D O I
10.1016/S0022-5347(17)35995-5
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
A total of 214 patients underwent implantation of the AMS 700 penile prosthesis with the controlled expansion cylinders. Followup ranged from 9 to 86 months, with a mean of 56 months. Life table analysis to calculate the probability of the cylinder and the device as a whole surviving 6 years revealed 97% for the cylinder and 90% for the device as a whole. No aneurysmal dilatation of the cylinders occurred. The incidence of cylinder leak was 0.7%. The infection rate in patients undergoing primary implantation of the device was 0.5% compared to 6.6% in patients undergoing revision surgery. The controlled expansion cylinders have significantly reduced the incidence of mechanical failures and, thus, the need for reoperation. As a consequence of the decreased need for reoperation the incidence of device infection has also decreased
引用
收藏
页码:46 / 48
页数:3
相关论文
共 13 条
[1]   EXPERIENCE WITH INFLATABLE PENILE PROSTHESIS [J].
FISHMAN, IJ ;
SCOTT, FB ;
LIGHT, JK .
UROLOGY, 1984, 23 :86-92
[2]   THE INFLATABLE PENILE PROSTHESIS - CLINICAL-EXPERIENCE WITH A NEW CONTROLLED EXPANSION CYLINDER [J].
FURLOW, WL ;
MOTLEY, RC .
JOURNAL OF UROLOGY, 1988, 139 (05) :945-946
[3]   ESTIMATING SURVIVAL FUNCTIONS FROM LIFE TABLE [J].
GEHAN, EA .
JOURNAL OF CHRONIC DISEASES, 1969, 21 (9-10) :629-+
[4]   SCOTT INFLATABLE PENILE PROSTHESIS - EVALUATION OF MECHANICAL SURVIVAL IN THE SERIES-700 MODEL [J].
GREGORY, JG ;
PURCELL, MH .
JOURNAL OF UROLOGY, 1987, 137 (04) :676-677
[5]   LONG-TERM EVALUATION OF THE INFLATABLE PENILE PROSTHESIS [J].
JOSEPH, DB ;
BRUSKEWITZ, RC ;
BENSON, RC .
JOURNAL OF UROLOGY, 1984, 131 (04) :670-673
[6]  
KNOLL LD, 1990, UROLOGY, V36, P502
[7]  
LIGHT JK, 1986, UROL SURG, V38, P1
[8]   RELIABILITY OF AMS M700 INFLATABLE PENILE PROSTHESIS [J].
MALLOY, TR ;
WEIN, AJ ;
CARPINIELLO, VL .
UROLOGY, 1986, 28 (05) :385-387
[9]   CLINICAL-EXPERIENCE WITH SCOTT INFLATABLE PENILE PROSTHESIS IN 150 PATIENTS [J].
MERRILL, DC .
UROLOGY, 1983, 22 (04) :371-375
[10]   USE OF CX CYLINDERS IN ASSOCIATION WITH AMS700 INFLATABLE PENILE PROSTHESIS [J].
MULCAHY, JJ .
JOURNAL OF UROLOGY, 1988, 140 (06) :1420-1421